RayStation 11B SPT1
by RaySearch Laboratories AB (PUBL)
RayStation 11B SPT1 is a radiation therapy software (FDA Class II; FDA product code MUJ, System, Planning, Radiation Therapy Treatment) made by RaySearch Laboratories AB (PUBL), cleared under FDA 510(k) K220141 on 15 Apr 2022 and listed in FDA GUDID since 12 May 2023 (UDI-DI 07350002010570), active as of 23 Oct 2026. 6 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
5 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number07350002010570 | Date12 May 2023 | StatusListed |
| United States | FDA 510(k)SESE | NumberK220141 | Date15 Apr 2022 | StatusActive |
Official records
- AccessGUDID record 07350002010570
- FDA 510(k) database K220141
From the FDA record
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA 510(k) K220141 decided 15 Apr 2022 Active
- FDA GUDID UDI-DI 07350002010570 listed 12 May 2023 Listed
Recalls
-
- Ongoing Z-1104-2026
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
-
- Ongoing Z-2492-2025
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
-
- Ongoing Z-1190-2025
Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
-
- Ongoing Z-1749-2024
Potential for reported SSD to be too high.
-
- Ongoing Z-1482-2024
A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.
-
- Terminated Z-1299-2022
This notice concerns two issues found related to the display of Linear Energy Transfer (LET) in RayStation 11B including some service packs. First, when using a dose threshold for an evaluation dose, LET display might be misleading. Second, a displayed beam-specific LET distribution can sometimes be out of sync with the selected beam.
Timeline
- Cleared FDA 510(k) K220141
- Recall initiated FDA recalls Z-1299-2022
- Listed FDA GUDID 07350002010570
- Recall initiated FDA recalls Z-1482-2024
- Recall initiated FDA recalls Z-1749-2024
- Recall initiated FDA recalls Z-1190-2025
- Recall initiated FDA recalls Z-2492-2025
- Recall initiated FDA recalls Z-1104-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- RayStation
- Generic name
- Radiation therapy software
- Model / reference
- 11B SPT1
- Primary DI / UDI-DI
- 07350002010570
- 510(k) Number
- FDA product code
- MUJ
- GMDN term
- Radiation therapy software
- Device class
- Class II
- Regulation number
- 892.5050
- GUDID distribution status
- In Commercial Distribution
Similar devices in this category
Other devices filed under Radiation therapy software.
- FilmQA Pro 7.0 N/A by Ashland Specialty Ingredients, G.P.
- VERIQA software S070031 by PTW-Freiburg Physikalisch-Technische Werkstätten Dr. Pychlau GmbH
- ART-Plan v.1.10.0 by Therapanacea Sas
- Standard Imaging IMSure QA Software v3 91326 by STANDARD IMAGING INC.
- FilmQA Pro 8.0 987957 by Ashland Specialty Ingredients, G.P.
- Vitesse 5.0 by Varian Medical Systems, Inc.