RayStation 11B
by RaySearch Laboratories AB (PUBL)
510(k) K220141
RayStation 11B is a medical device (FDA Class II; FDA product code MUJ, System, Planning, Radiation Therapy Treatment) made by RaySearch Laboratories AB (PUBL), cleared under FDA 510(k) K220141 on 15 Apr 2022, active as of 22 Oct 2026. 6 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
5 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK220141 | Date15 Apr 2022 | StatusActive |
Official records
- FDA 510(k) database K220141
From the FDA record
- Received by the FDA
- 18 Jan 2022
- FDA decision
- 15 Apr 2022 (Substantially Equivalent)
- FDA review time
- 87 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Ongoing Z-1104-2026
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
-
- Ongoing Z-2492-2025
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
-
- Ongoing Z-1190-2025
Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
-
- Ongoing Z-1749-2024
Potential for reported SSD to be too high.
-
- Ongoing Z-1482-2024
A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.
-
- Terminated Z-1299-2022
This notice concerns two issues found related to the display of Linear Energy Transfer (LET) in RayStation 11B including some service packs. First, when using a dose threshold for an evaluation dose, LET display might be misleading. Second, a displayed beam-specific LET distribution can sometimes be out of sync with the selected beam.
Timeline
- Cleared FDA 510(k) K220141
- Recall initiated FDA recalls Z-1299-2022
- Recall initiated FDA recalls Z-1482-2024
- Recall initiated FDA recalls Z-1749-2024
- Recall initiated FDA recalls Z-1190-2025
- Recall initiated FDA recalls Z-2492-2025
- Recall initiated FDA recalls Z-1104-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- MUJ
- Device class
- Class II
- Regulation
- 21 CFR Part 892.5050
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Devices cleared under K220141
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| RayStation 11B SPT1 | 11B SPT1 | 07350002010570 | — | On the market |
Similar devices
Other devices filed under FDA product code MUJ (System, Planning, Radiation Therapy Treatment).
- RT Elements (4.5); (Elements) Multiple Brain Mets SRS; (Elements) Cranial SRS; (Elements) Spine SRS; (Elements) Cranial SRS w/ Cones; (Elements) RT Planning Platform; (Elements) Dose Review; (Elements) Retreatment Review; Elements Segmentation [Cranial, Basal Ganglia, Head & Neck, Pelvic, Spine, Thoracic & Spine, Extracranial] RT; Elements AI Tumor Segmentation RT; Elements SmartBrush [Angio, Spine] RT; Elements Object Management RT by Brainlab AG
- BRACHYVISION 6.0 by Varian Assoc., Inc.
- PINNACLE3 SMARTARC by Philips Medical Systems
- Plan AI 1057 by Sun Nuclear Corporation
- syngo.via RT Image Suite 11546418 by Siemens Healthcare GmbH
- SEGMENTATION ASSISTANT FOR PROSTATE-AUTO CONTOURING by Omisa Inc. (Segasist Technologies)
- RAYSTATION by Raysearch Laboratories AB
- ACQPLAN 5.0 by Philips Medical Systems
- RAY STATION by Raysearch Laboratories AB
- MONACO by Impac Medical Systems, Inc.