Skip to content

RayStation 11B

by RaySearch Laboratories AB (PUBL)

510(k) K220141

RayStation 11B is a medical device (FDA Class II; FDA product code MUJ, System, Planning, Radiation Therapy Treatment) made by RaySearch Laboratories AB (PUBL), cleared under FDA 510(k) K220141 on 15 Apr 2022, active as of 22 Oct 2026. 6 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

5 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK220141 Date15 Apr 2022ClearedStatusActive

Official records

From the FDA record

Received by the FDA
18 Jan 2022
FDA decision
15 Apr 2022 (Substantially Equivalent)
FDA review time
87 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K220141 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

6 recalls · 5 ongoing · FDA

    • Ongoing Z-1104-2026 FDA Class II initiated 28 Nov 2025

    Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

    • Ongoing Z-2492-2025 FDA Class II initiated 8 Aug 2025

    DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.

    • Ongoing Z-1190-2025 FDA Class II initiated 16 Jan 2025

    Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.

    • Ongoing Z-1749-2024 FDA Class II initiated 28 Mar 2024

    Potential for reported SSD to be too high.

    • Ongoing Z-1482-2024 FDA Class II initiated 23 Feb 2024

    A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.

    • Terminated Z-1299-2022 FDA Class II initiated 10 Jun 2022

    This notice concerns two issues found related to the display of Linear Energy Transfer (LET) in RayStation 11B including some service packs. First, when using a dose threshold for an evaluation dose, LET display might be misleading. Second, a displayed beam-specific LET distribution can sometimes be out of sync with the selected beam.

Timeline

7 dated events

  1. Cleared FDA 510(k) K220141
  2. Recall initiated FDA recalls Z-1299-2022
  3. Recall initiated FDA recalls Z-1482-2024
  4. Recall initiated FDA recalls Z-1749-2024
  5. Recall initiated FDA recalls Z-1190-2025
  6. Recall initiated FDA recalls Z-2492-2025
  7. Recall initiated FDA recalls Z-1104-2026

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
MUJ
Device class
Class II
Regulation
21 CFR Part 892.5050
Advisory committee
Radiology
FDA decision
SESE Substantially Equivalent

Devices cleared under K220141

1 device · FDA GUDID

Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.

Devices cleared under K220141
UDI-DISize and differences
RayStation 11B SPT1
RaySearch Laboratories AB (PUBL)
11B SPT107350002010570 —On the market

Similar devices

Other devices filed under FDA product code MUJ (System, Planning, Radiation Therapy Treatment).

Other devices from RaySearch Laboratories AB (PUBL)

See all devices from RaySearch Laboratories AB (PUBL) →