Raystation
by Raysearch Laboratories AB 3 variants
510(k) K120387
Raystation is a Class II device made by Raysearch Laboratories AB, cleared under FDA 510(k) K120387 on 6 Apr 2012, active as of 22 Oct 2026. 6 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK120387 | Date6 Apr 2012 | StatusActive |
Official records
- FDA 510(k) database K120387
Recalls
-
- Terminated Z-1679-2015
RaySearch Laboratories AB became aware of an issue with the dose calculation for some imported VMAT plans where controls points are defined with wide gantry angle spacing.
-
- Terminated Z-1481-2014
RaySearch Laboratories has recalled "RaySearch Ray Station 2.0, 2.5, 3.0, 3.5 and 4.0" software due to dose miscalculations for bolus/external/fixation support structures on CT-image.
-
- Terminated Z-1307-2014
Potential for dose errors due to software program errors.
-
- Terminated Z-0281-2014
Ray Search Laboratories found an issue with the dose calculation for Elekta motorized wedges in Ray Station 3.0, 3.5 and 4.0. The error presents when a wedge is used for a beam and the wedge angle differs from 60 degrees.
-
- Terminated Z-1419-2013
Under some circumstances, imported CT, MR and PET images are offset from their true positions by one pixel in one or two directions. Users must be aware of this information to avoid incorrect dose calculations during treatment planning.
-
- Terminated Z-2324-2012
The firm decided to recall when an issue was found with the dose calculation in RayStation 2.5, Raystation 2.5.1. The problem has not caused any patient mistreatment or other incidents. However, the user must be aware of this information to avoid incorrect dose calculations during treatment planning. The dose calculation is not correctly normalized in the unusual case of preventing the optimizatio
Timeline
- Cleared FDA 510(k) K120387
- Recall initiated FDA recalls Z-2324-2012
- Recall initiated FDA recalls Z-1419-2013
- Recall initiated FDA recalls Z-0281-2014
- Recall initiated FDA recalls Z-1307-2014
- Recall initiated FDA recalls Z-1481-2014
- Recall initiated FDA recalls Z-1679-2015
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- MUJ
- Device class
- Class II
- Regulation
- 21 CFR Part 892.5050
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code MUJ (System, Planning, Radiation Therapy Treatment).
- ACQPLAN by Philips Medical Systems (Cleveland), Inc.
- Multimodality Simulation Workspace (MM Sim) (v1.0.0) by Philips Medical Systems Nederland B.V.
- ClearCalc Model RADCA V2 by Radformation, Inc.
- 4D INTEGRATED TREATMENT CONSOLE by Varian Medical Systems, Inc.
- Panther TPS by Prowess, Inc.
- ART-Plan v.1.10.0 by Therapanacea Sas
- HI-ART SYSTEM by Tomo Therapy, Inc.
- MMS THERPACPLUS B3DTUI MODEL NUMBER 6.7 by Multimedia Medical Systems, Inc.
- PLATO SRS STEREOTACTIC RADIOSURGERY PLANNING SYSTEM, MODEL 2.0, PLATO IFS IMAGE FUSION SOFTWARE, MODEL 2.1 by Nucletron Corp.
- ITP, MODEL 1.0 by Nucletron Corp.