Raystation
by Raysearch Laboratories AB 3 variants
510(k) K130617
Raystation is a Class II device made by Raysearch Laboratories AB, cleared under FDA 510(k) K130617 on 10 May 2013, active as of 22 Oct 2026. 6 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK130617 | Date10 May 2013 | StatusActive |
Official records
- FDA 510(k) database K130617
Recalls
-
- Terminated Z-1679-2015
RaySearch Laboratories AB became aware of an issue with the dose calculation for some imported VMAT plans where controls points are defined with wide gantry angle spacing.
-
- Terminated Z-1806-2014
This notice concerns a problem with the computation of ROI voxel volumes for cases where the CT has variable slice spacing. The dose grid volumetric representation of an ROI can be wrong. This affects all dose-volume properties for such ROIs, including DVH, dose statistics, clinical goals and constraints or objective functions. Furthermore, if material override ROIs are used, or if the external is not represented by contours, the bug can trigger an error in the 3D dose distribution.
-
- Terminated Z-1481-2014
RaySearch Laboratories has recalled "RaySearch Ray Station 2.0, 2.5, 3.0, 3.5 and 4.0" software due to dose miscalculations for bolus/external/fixation support structures on CT-image.
-
- Terminated Z-1307-2014
Potential for dose errors due to software program errors.
-
- Terminated Z-1308-2014
RaySearch Laboratories has recalled RaySearch Ray Station 4.0 software due to clinical dose calculation errors during radiation therapy.
-
- Terminated Z-0281-2014
Ray Search Laboratories found an issue with the dose calculation for Elekta motorized wedges in Ray Station 3.0, 3.5 and 4.0. The error presents when a wedge is used for a beam and the wedge angle differs from 60 degrees.
Timeline
- Cleared FDA 510(k) K130617
- Recall initiated FDA recalls Z-0281-2014
- Recall initiated FDA recalls Z-1307-2014
- Recall initiated FDA recalls Z-1308-2014
- Recall initiated FDA recalls Z-1481-2014
- Recall initiated FDA recalls Z-1806-2014
- Recall initiated FDA recalls Z-1679-2015
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- MUJ
- Device class
- Class II
- Regulation
- 21 CFR Part 892.5050
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code MUJ (System, Planning, Radiation Therapy Treatment).
- P.D by .Decimal, LLC
- Sensus TPS Workstation by Sensus Healthcare
- KONRAD, MODEL V.2.0 by Mrc Systems GmbH
- 4D Integrated Treatment Console by Varian Medical Systems, Inc.
- TRAPIS (TRANSPERINEAL PROSTATE IMPLANT SIMULATOR), VERSION 1.5 by Kps & Assoc., Inc.
- ERGO++ VERSION 1.7 by 3D Line Medical Systems S.R.L.
- INTEGRA RADIONICS XKNIFE RT 4 by Integra Radionics, Inc.
- CROSSPLAN / ACCUSOFT-XL V.4.04 by Direx Systems Corp.
- RayStation 9B SP1 by RaySearch Laboratories AB (PUBL)
- VITESSE 3.0 by Varian Medical Systems, Oncology Systems