RayStation
by RaySearch Laboratories AB (PUBL) 4 variants
510(k) K190387
RayStation is a Class II device made by RaySearch Laboratories AB (PUBL), cleared under FDA 510(k) K190387 on 19 Jun 2019, active as of 22 Oct 2026. 32 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
8 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK190387 | Date19 Jun 2019 | StatusActive |
Official records
- FDA 510(k) database K190387
Recalls
-
- Ongoing Z-2488-2025
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
-
- Ongoing Z-1186-2025
Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
-
- Ongoing Z-1743-2024
- Ongoing Z-1744-2024
Potential for reported SSD to be too high.
-
- Ongoing Z-1482-2024
A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.
-
- Ongoing Z-1281-2023
It is possible to set a non-zero collimator angle in the Virtual Simulation module. The angle will be correct in the user interface but if the plan is DICOM exported by using the VSIM Export button, the angle will not be correct in the exported RT plan. In the exported RT plan, the DICOM attribute Beam Limiting Device Angle (300A,0120) is set to zero, regardless of the selected collimator angle.
-
- Ongoing Z-2426-2020
It is not obvious that the manual import of a treatment record into RayCare PACS does not automatically import the same data into RayTreat. In addition, there is an issue found with the display of planned and recorded Gantry Angle in the delivered beam list in RayTreat.
-
- Ongoing Z-1598-2020
Due to physical limitations of the treatment machine, the maximum field opening for Siemens Virtual Wedge fields recommended by the LINAC vendor will be more restricted in the wedged direction than the maximum field opening without a wedge. Delivering a larger field opening with a virtual wedge will not result in a wedged field shape over the entire field. Instead, a high dose plateau region will be the result at the toe side of the field beyond the recommended field size.
-
- Terminated Z-2343-2020
- Terminated Z-2351-2020
- Terminated Z-2352-2020
- Terminated Z-2347-2020
- Terminated Z-2349-2020
- Terminated Z-2348-2020
- Terminated Z-2339-2020
- Terminated Z-2345-2020
- Terminated Z-2350-2020
- Terminated Z-2353-2020
- Terminated Z-2340-2020
- Terminated Z-2346-2020
- Terminated Z-2342-2020
- Terminated Z-2341-2020
- Terminated Z-2344-2020
Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
-
- Terminated Z-1465-2020
- Terminated Z-1463-2020
- Terminated Z-1464-2020
- Terminated Z-1468-2020
- Terminated Z-1462-2020
- Terminated Z-1467-2020
- Terminated Z-1466-2020
There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.
-
- Terminated Z-0866-2020
Three issues found: i) The Map ROI options in the ROI list in the Structure Definition module may generate unintended ROI geometries ii) Elekta guard leaf behavior. There is an interoperability issue with Elekta regarding setting of guard leaves. iii)SSD, when intended as source-to-surface distance, it sometimes gives source-to-skin distance.To the best of our knowledge, these issues have not caused any patient mistreatment or other incidents. However, the user must be aware of the following information to avoid incorrect dose calculations during treatment planning.
-
- Terminated Z-0736-2020
Editing a static arc beam may unexpectedly set the MU of the beam to its initial default value of 200 MU. If the bug is triggered, the dose is invalidated and needs to be recalculated before the plan can be approved or exported.
Timeline
- Recall initiated FDA recalls Z-2342-2020
- Recall initiated FDA recalls Z-2341-2020
- Recall initiated FDA recalls Z-2344-2020
- Recall initiated FDA recalls Z-2426-2020
- Recall initiated FDA recalls Z-1281-2023
- Recall initiated FDA recalls Z-1482-2024
- Recall initiated FDA recalls Z-1743-2024
- Recall initiated FDA recalls Z-1744-2024
- Recall initiated FDA recalls Z-1186-2025
- Recall initiated FDA recalls Z-2488-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- MUJ
- Device class
- Class II
- Regulation
- 21 CFR Part 892.5050
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Devices cleared under K190387
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| RayStation 8.1 | 8.1 | 07350002010129 | — | On the market |
| RayStation 9 | 9 | 07350002010174 | — | On the market |
Similar devices
Other devices filed under FDA product code MUJ (System, Planning, Radiation Therapy Treatment).
- RT Elements, Cranial SRS, Spine SRS, Multiple Brain Mets SRS, RT QA, Adaptive Hybrid Surgery Analysis, Dose Review by Brainlab AG
- ECLIPSE WITH AAA by Varian Medical Systems, Inc
- VIEWRAY TREATMENT PLANNING AND DELIVERY SYSTEM by VIEWRAY INCORPORATED
- DCM 1.0 by Mds Nordion, Inc.
- Monaco RTP System (6.3) by ELEKTA SOLUTIONS AB
- RAYSTATION VERSION 1.0 by Raysearch Laboratories AB
- RADIONICS XKNIFE RT 3 WITH NON STEREOTACTIC MODULE by Radionics, A Division of Tyco Healthcare Group LP
- syngo.via RT Image Suite by SIEMENS MEDICAL SOLUTIONS USA, INC.
- CYBERKNIFE SYSTEM FOR STEREOTACTIC RADIOSURGERY/RADIOTHERAPY by Accuray Incorporated
- CORVUS Radiation Therapy Planning System by Best Nomos