ROSA ONE Brain application
510(k) K200511
Active
510(k)
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK200511 | Date29 May 2020 | StatusActive |
Recalls
-
- Ongoing Z-0118-2022
The firm has become aware of a software anomaly affecting ROSA One 3.1 Brain application which led to the inaccurate placement of an electrode during surgery. The firm has received 3 global complaints related to the issue. An incorrect trajectory could result in serious injury or death if undetected during surgery.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- HAW
- Device Class
- Class II
- Regulation
- 21 CFR Part 882.4560
- Advisory Committee
- Neurology
- FDA Decision
- SESE Substantially Equivalent