Skip to content

Saccadometer Advanced

by Ober Consulting Sp. Z.O.O.

Saccadometer Advanced is an eye movement analysis system (FDA Class II; FDA product code HLL, Monitor, Eye Movement) made by Ober Consulting Sp. Z.O.O., cleared under FDA 510(k) K152890 on 11 Aug 2016 and listed in FDA GUDID since 26 Jan 2017 (UDI-DI 05906874392027), active as of 22 Oct 2026.

Active United States. Also listed in GUDID (US).

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None found

in data synced 22 Oct 2026

Documents

510(k) K152890

FDA submission, from GUDID

Trust

Unverified

Source GUDID, synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number05906874392027 UDI-DI Date26 Jan 2017ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE from GUDID Substantially Equivalent · Traditional NumberK152890 Date11 Aug 2016DecidedStatusActive

Official records

From the FDA record

Device description
The Saccadometer Adavnced system measures eye movements in horizontal axis with temporal 1kHz and spatial resolution < 5 arc min. The system incorporates a built in visual target presentation using miniature laser projectors mounted on the sensor forehead plate. Saccadometer Advanced is equipped with three red laser projectors with 10º separation and green laser located centrally over the red. Eighteen experiments are included in embedded software. The information about the eye movement is stored in a small, self-contained control unit, and can be downloaded for computer analysis at a later stage. Information about eye movement is acquired by analyzing the infrared light reflected from eye orbit. The left and right eye orbit are illuminated by two infrared LEDs, providing eye surface irradiance below 1mW/cm 2 (operating condition). Due to the conjugacy and synchronicity of the saccadic eye movements, the rotation of left and right eyes, can be added and averaged. The inner canthi of the left and right eyes are illuminated with the low intensity IR. The difference between the amounts of IR reflected back from the eye surfaces toward the photodetecting circuitry, carries the information about the eye position changes. Two eye landmarks are taking part in generation of the eye movement signal: the corneal bulge and the limbus (the border between darker iris and white sclera). Main contributor to the eye movement signal is the corneal bulge. Being the relative eye position measurement system, it requires to define the initial eye position. It means that at the beginning of every measurement, a steady eye fixation point needs to be provided. In Saccadometer the central fixation target serves this purpose.
MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
No latex statement on the label
Sterility
Supplied non-sterile

From the GUDID record (AccessGUDID), as read on 22 Oct 2026, as the labeller entered it.

Across registries

2 registries list this device

  • FDA 510(k) K152890 decided 11 Aug 2016 Active
  • FDA GUDID UDI-DI 05906874392027 listed 26 Jan 2017 Listed

Timeline

2 dated events

  1. Cleared FDA 510(k) K152890
  2. Listed FDA GUDID 05906874392027

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

GUDID

Trade name
Saccadometer
Generic name
Eye movement analysis system
Model / reference
ADVANCED
Primary DI / UDI-DI
05906874392027
510(k) Number
FDA product code
HLL
GMDN term
Eye movement analysis system
Device class
Class II
Regulation number
886.1510
GUDID distribution status
In Commercial Distribution

Similar devices

Other devices filed under FDA product code HLL (Monitor, Eye Movement).

Other devices from Ober Consulting Sp. Z.O.O.

Properties

  • Prescription use
  • Serial number
  • Manufacturing date
Other properties (12, all answered No)