Saccadometer Plus
Saccadometer Plus is an eye movement analysis system (FDA Class II; FDA product code HLL, Monitor, Eye Movement) made by Ober Consulting Sp. Z.O.O., cleared under FDA 510(k) K152890 on 11 Aug 2016 and listed in FDA GUDID since 26 Jan 2017 (UDI-DI 05906874392010), active as of 22 Oct 2026.
Active
510(k)
Valid until
No expiry
FDA recalls
None found
Documents
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number05906874392010 | Date26 Jan 2017 | StatusListed |
| United States | FDA 510(k)SESE | NumberK152890 | Date11 Aug 2016 | StatusActive |
Official records
- AccessGUDID record 05906874392010
- FDA 510(k) database K152890
From the FDA record
- Device description
- The Saccadometer Plus system measures eye movements in horizontal axis with temporal 1kHz and spatial resolution < 5 arc min. The system incorporates a built in visual target presentation using miniature laser projectors mounted on the sensor forehead plate. Saccadometer Plus is equipped with three red laser projectors with 10º separation, and 6 experiments are included in embedded software. The information about the eye movement is stored in a small, self-contained control unit, and can be downloaded for computer analysis at a later stage. Information about eye movement is acquired by analyzing the infrared light reflected from eye orbit. The left and right eye orbit are illuminated by two infrared LEDs, providing eye surface irradiance below 1mW/cm 2 (operating condition). Due to the conjugacy and synchronicity of the saccadic eye movements, the rotation of left and right eyes, can be added and averaged. The inner canthi of the left and right eyes are illuminated with the low intensity IR. The difference between the amounts of IR reflected back from the eye surfaces toward the photodetecting circuitry, carries the information about the eye position changes. Two eye landmarks are taking part in generation of the eye movement signal: the corneal bulge and the limbus (the border between darker iris and white sclera). Main contributor to the eye movement signal is the corneal bulge. Being the relative eye position measurement system, it requires to define the initial eye position. It means that at the beginning of every measurement, a steady eye fixation point needs to be provided. In Saccadometer the central fixation target serves this purpose.
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA 510(k) K152890 decided 11 Aug 2016 Active
- FDA GUDID UDI-DI 05906874392010 listed 26 Jan 2017 Listed
Timeline
- Cleared FDA 510(k) K152890
- Listed FDA GUDID 05906874392010
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Saccadometer
- Generic name
- Eye movement analysis system
- Model / reference
- PLUS
- Primary DI / UDI-DI
- 05906874392010
- 510(k) Number
- FDA product code
- HLL
- GMDN term
- Eye movement analysis system
- Device class
- Class II
- Regulation number
- 886.1510
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code HLL (Monitor, Eye Movement).