Sarn Centrifugal Pump
by Terumo Cardiovascular Systems Corporation
510(k) K112229
Sarn Centrifugal Pump is a medical device (FDA Class II; FDA product code KFM, Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type) made by Terumo Cardiovascular Systems Corporation, cleared under FDA 510(k) K112229 on 9 Dec 2011, active as of 22 Oct 2026. 4 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK112229 | Date9 Dec 2011 | StatusActive |
Official records
- FDA 510(k) database K112229
From the FDA record
- Received by the FDA
- 3 Aug 2011
- FDA decision
- 9 Dec 2011 (Substantially Equivalent)
- FDA review time
- 128 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Terminated Z-1235-2017
Terumo Cardiovascular Systems Corporation has received complaints indicating that the Sarns Centrifugal Pump (disposable pump heads) have exhibited leaks. The leaks were detected during priming of the bypass circuit and during CPB procedures. In each of the reported incidents, there were no reports of patient injury as a result of the centrifugal pump leaks. Terumo Cardiovascular System's investigation into the incident has identified the cause of the leaks to be cracks in both the Top Housing and Magnet Back Housing of the pump. When the cracks cause the housing to become compromised, it is possible for fluid to leak from the pump head.
-
- Terminated Z-2028-2014
- Terminated Z-2029-2014
- Terminated Z-2030-2014
During set-up and priming of the bypass circuit, leaks were detected on the centrifugal pumphead.
Timeline
- Cleared FDA 510(k) K112229
- Recall initiated FDA recalls Z-2028-2014
- Recall initiated FDA recalls Z-2029-2014
- Recall initiated FDA recalls Z-2030-2014
- Recall initiated FDA recalls Z-1235-2017
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- KFM
- FDA applicant
- Terumo Cardiovascular Systems
- Device class
- Class II
- Regulation
- 21 CFR Part 870.4360
- Advisory committee
- Cardiovascular
- FDA decision
- SESE Substantially Equivalent
Devices cleared under K112229
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| Sarns™ Centrifugal Pump Adapter 824043 | 824043 | 00699753450417 | On the market | |
| Sarns™ Centrifugal Pump with X-Coating™ 164275X | 164275X | 00699753450325 | On the market | |
| Sarns™ Centrifugal Pump with X-Coating™ CV-164275X | CV-164275X | 00699753450349 | On the market | |
| Sarns™ Centrifugal Pump without X-Coating™ 164275 | 164275 | 00699753002333 | On the market | |
| Sarns™ Centrifugal Pump without X-Coating™ CV-164275 | CV-164275 | 00699753450332 | On the market |
Similar devices
Other devices filed under FDA product code KFM (Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type).
- JOSTRA MECC SYSTEM by Jostra AG
- ECC.O SYSTEM by Dideco S.R.L.
- IBC FLOPUMP WITH GBS COATING by Gish Biomedical, Inc.
- ROTAFLOW CENTRIFUGAL PUMP SYSTEM by Jostra Medizintechnik AG
- NOVOSCI READY SYSTEM, MODEL RS-200 by NOVOSCI CORP.
- Rotaflow Centrifugal Pump With Softline Coating, Model BO-RF-32 by Maquet Cardiopulmonary AG
- IDEAL MIMESYS SYSTEM WITH INTEGRATED VENOUS AIR REMOVAL, CENTRIFUGAL BLOOD PUMP, PUMP BRACKET, ADULT MEMBRANE OXYGENATOR by Dideco S.P.A.
- CAPIOX SP PUMP HEAD by TERUMO MEDICAL CORPORATION
- CENTRIMED SYSTEM 1 by Centrimed Corp.
- HEPARIN COATED MINIATURE CENTRIFUGAL BYPASS PUMP SYSTEM;MODEL #CP-1H by A-Med Systems, Inc.
Other devices from Terumo Cardiovascular Systems Corporation
See all devices from Terumo Cardiovascular Systems Corporation →