Skip to content

Sarn Centrifugal Pump

by Terumo Cardiovascular Systems Corporation

510(k) K112229

Sarn Centrifugal Pump is a medical device (FDA Class II; FDA product code KFM, Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type) made by Terumo Cardiovascular Systems Corporation, cleared under FDA 510(k) K112229 on 9 Dec 2011, active as of 22 Oct 2026. 4 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

4 closed

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK112229 Date9 Dec 2011ClearedStatusActive

Official records

From the FDA record

Received by the FDA
3 Aug 2011
FDA decision
9 Dec 2011 (Substantially Equivalent)
FDA review time
128 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K112229 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

4 recalls · FDA

    • Terminated Z-1235-2017 FDA Class II initiated 27 May 2015

    Terumo Cardiovascular Systems Corporation has received complaints indicating that the Sarns Centrifugal Pump (disposable pump heads) have exhibited leaks. The leaks were detected during priming of the bypass circuit and during CPB procedures. In each of the reported incidents, there were no reports of patient injury as a result of the centrifugal pump leaks. Terumo Cardiovascular System's investigation into the incident has identified the cause of the leaks to be cracks in both the Top Housing and Magnet Back Housing of the pump. When the cracks cause the housing to become compromised, it is possible for fluid to leak from the pump head.

  • 3 recalls, same reason

    • Terminated Z-2028-2014 FDA Class II initiated 14 Apr 2014
    • Terminated Z-2029-2014 FDA Class II initiated 14 Apr 2014
    • Terminated Z-2030-2014 FDA Class II initiated 14 Apr 2014

    During set-up and priming of the bypass circuit, leaks were detected on the centrifugal pumphead.

Timeline

5 dated events

  1. Cleared FDA 510(k) K112229
  2. Recall initiated FDA recalls Z-2028-2014
  3. Recall initiated FDA recalls Z-2029-2014
  4. Recall initiated FDA recalls Z-2030-2014
  5. Recall initiated FDA recalls Z-1235-2017

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
KFM
FDA applicant
Terumo Cardiovascular Systems
Device class
Class II
Regulation
21 CFR Part 870.4360
Advisory committee
Cardiovascular
FDA decision
SESE Substantially Equivalent

Devices cleared under K112229

5 devices · FDA GUDID

Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.

Devices cleared under K112229
UDI-DISize and differences
Sarns™ Centrifugal Pump Adapter 824043
Terumo Cardiovascular Systems Corporation
82404300699753450417
Single use: no
On the market
Sarns™ Centrifugal Pump with X-Coating™ 164275X
Terumo Cardiovascular Systems Corporation
164275X00699753450325
Single use: yes
On the market
Sarns™ Centrifugal Pump with X-Coating™ CV-164275X
Terumo Cardiovascular Systems Corporation
CV-164275X00699753450349
Single use: yes
On the market
Sarns™ Centrifugal Pump without X-Coating™ 164275
Terumo Cardiovascular Systems Corporation
16427500699753002333
Single use: yes
On the market
Sarns™ Centrifugal Pump without X-Coating™ CV-164275
Terumo Cardiovascular Systems Corporation
CV-16427500699753450332
Single use: yes
On the market

Similar devices

Other devices filed under FDA product code KFM (Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type).

Other devices from Terumo Cardiovascular Systems Corporation

See all devices from Terumo Cardiovascular Systems Corporation →