by Howmedica Osteonics Corp.

Scorpio 82-3-0708 is a Class II device made by Howmedica Osteonics Corp., cleared under FDA 510(k) K261898 on 6 Aug 2026 (2 earlier submissions) and listed in FDA GUDID since 24 Sep 2015 (UDI-DI 04546540364418), active as of 23 Oct 2026. 12 FDA recalls are linked to it.

Active United States. Also listed in GUDID (US).

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

3 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source GUDID, synced 23 Oct 2026.

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE from GUDID Substantially Equivalent · Traditional NumberK261898 Date6 Aug 2026DecidedStatusActive
United StatesGUDID listing Number04546540364418 UDI-DI Date24 Sep 2015ListedStatusListed GUDID: in commercial distribution

Official records

Across registries

2 registries list this device

  • FDA 510(k) K261898 decided 6 Aug 2026 Active· +2 more in the status table
  • FDA GUDID UDI-DI 04546540364418 listed 24 Sep 2015 Listed

Recalls

13 recalls · 3 ongoing · FDA

  • 2 recalls, same reason

    • Ongoing Z-3206-2026 FDA Class II initiated 30 Jul 2026 on shared 510(k) K243817
    • Ongoing Z-3197-2026 FDA Class II initiated 30 Jul 2026

    Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

    • Ongoing Z-0899-2024 FDA Class II initiated 1 Dec 2023

    Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

    • Terminated Z-1634-2015 FDA Class II initiated 15 Aug 2014

    The peel strength of the inner blister may have been below internal validated requirements. The strength of the seal is directly related to long term shelf life of the sterile barrier.

  • 8 recalls, same reason

    • Terminated Z-0471-2010 FDA Unclassified (pre-2012) initiated 2 Oct 2009
    • Terminated Z-0477-2010 FDA Unclassified (pre-2012) initiated 2 Oct 2009
    • Terminated Z-0475-2010 FDA Unclassified (pre-2012) initiated 2 Oct 2009
    • Terminated Z-0474-2010 FDA Unclassified (pre-2012) initiated 2 Oct 2009
    • Terminated Z-0473-2010 FDA Unclassified (pre-2012) initiated 2 Oct 2009
    • Terminated Z-0476-2010 FDA Unclassified (pre-2012) initiated 2 Oct 2009
    • Terminated Z-0470-2010 FDA Unclassified (pre-2012) initiated 2 Oct 2009
    • Terminated Z-0472-2010 FDA Unclassified (pre-2012) initiated 2 Oct 2009

    It was determined that a change made in the dimensions of the collar component of the packaging resulted in the inner blister material being overstretched causing a potential hole in the sterile barrier system.

    • Terminated Z-1141-2010 FDA Unclassified (pre-2012) initiated 20 Aug 2007

    A box labeled as a Scorpio NRG Femoral Size 7 Right, may actually contain a Scorpio PS Femoral Size 5 Right.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

17 dated events

  1. Recall initiated FDA recalls Z-0476-2010
  2. Recall initiated FDA recalls Z-0470-2010
  3. Recall initiated FDA recalls Z-0472-2010
  4. Recall initiated FDA recalls Z-1634-2015
  5. Listed FDA GUDID 04546540364418
  6. Recall initiated FDA recalls Z-0899-2024
  7. Cleared FDA 510(k) K243817
  8. Recall initiated FDA recalls Z-3206-2026
  9. Recall initiated FDA recalls Z-3197-2026
  10. Cleared FDA 510(k) K261898

7 earlier events are in the sections above.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

GUDID

Trade name
Scorpio
Generic name
Tibial insert
Model / Ref / Catalog
82-3-0708
Primary DI / UDI-DI
04546540364418
FDA product code
JWH
GMDN term
Tibial insert
Device class
Class II
Regulation number
888.3560
GUDID distribution status
In Commercial Distribution

Similar devices in this category

Other devices filed under Tibial insert.

Other devices from Howmedica Osteonics Corp.

See all devices from Howmedica Osteonics Corp. →

Properties

  • Supplied sterile
  • Single use
  • Implantable
  • Prescription use
  • Lot / batch number
Other properties (10, all answered No)