Selenia® Dimensions® ASY-08483-2D-CN
by Hologic GUDID
Listed (no authorization)
Regulatory status
Recall ongoing. 3 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number15420045518230 | Date12 Mar 2026 | StatusIn distribution |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — GUDID listing Listed
Recalls
-
- Ongoing Z-2168-2026
Complaints have been received of systems developing loose, missing, or broken internal bolts over time.
-
- Ongoing Z-3248-2024
Identified an issue with Selenia Dimensions 1.12.0 and 3Dimensions 2.3.0 system software when used in combination with C-View software in which the quality of the final C-View synthesized 2D image is impacted and may appear blurry and result in misdiagnosis with the possibility being a false negative diagnosis
-
- Ongoing Z-1144-2023
C-Arm unexpected movement may cause blunt trauma should the tube arm impinge upon an individual
-
- Terminated Z-1343-2020
The recalling firm has received reports from the field of the C-arm on the Selenia Dimensions system seizing or dropping during field repairs and cleaning of the systems.
Specification
- Trade Name
- Selenia® Dimensions®
- Generic Name
- Stationary mammographic x-ray system, digital
- Model / Ref / Catalog
- ASY-08483-2D-CN
- Primary DI / UDI-DI
- 15420045518230
- PMA Number
- FDA Product Code
- OTE
- GMDN Term
- Stationary mammographic x-ray system, digital
- Commercial Distribution
- In Commercial Distribution