Skip to content

Somatom Spirit

by Siemens Medical Systems, Inc.

510(k) K042328

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

2 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK042328 Date14 Sep 2004ClearedStatusActive

Recalls

2 recalls · FDA

    • Terminated Z-1451-2018 FDA Class II initiated 15 Feb 2018

    to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

    • Terminated Z-1721-2014 FDA Class II initiated 21 Mar 2014

    The rubber damper may degrade and break, causing loosening of the motor mounting plates, which could lead to a series of events that could place the operator and patient at risk of injury.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
JAK
Device Class
Class II
Regulation
21 CFR Part 892.1750
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent

Other devices from Siemens Medical Systems USA, Inc.