Spacelabs Healthcare Xprezzon Beside Monitor(9913930)
510(k) K112962
Active
Valid until
No expiry
FDA recalls
3 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK112962 | Date2 Nov 2011 | StatusActive |
Recalls
-
- Ongoing Z-0169-2026
- Ongoing Z-0167-2026
- Ongoing Z-0168-2026
Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.
-
- Terminated Z-1144-2015
Spacelabs Healthcare XPREZZON Bedside Monitors, Model 91393, are recalled because the firm has received multiple reports of XPREZZON monitors failing to boot up or returning to factory default configuration settings following power on or reset.
-
- Terminated Z-0196-2013
Potential failure of Spacelabs Healthcare Touchscreen Display, Model 94267 where the alarm lights failed to illuminate upon installation.
-
- Terminated Z-2019-2012
There were reports that a certain combination of key presses will cause the module to disconnect from the Xprezzon bedside monitor, model 91393, and the patient parameters will drop from the display. The parameters return after power cycling the monitor.
-
- Terminated Z-1705-2012
The screw used to connect the external power supply to the Spacelabs Medical Xprezzon Bedside Monitor, Model 91393, may damage an internal capacitor. The damage may result in an inability of the monitor to power on, power off, or involuntarily shut down.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MHX
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.1025
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent