STar™ Drive M/E System ST-DS-ME-00443
by Fhc, Inc.
STar Drive M/E System ST-DS-ME-00443 is a neurological stereotactic surgery system (FDA Class II; FDA product code HAW, Neurological stereotaxic Instrument) made by Fhc, Inc., cleared under FDA 510(k) K092562 on 18 Sep 2009 and listed in FDA GUDID since 17 Sep 2016 (UDI-DI 00873263004432), active as of 22 Oct 2026. 9 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00873263004432 | Date17 Sep 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK092562 | Date18 Sep 2009 | StatusActive |
Official records
- AccessGUDID record 00873263004432
- FDA 510(k) database K092562
From the FDA record
- Device description
- STar Drive Motor/Encoder System
- MRI safety
- MR Unsafe
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile; sterilise before use (Moist Heat or Steam Sterilization)
Across registries
- FDA 510(k) K092562 decided 18 Sep 2009 Active
- FDA GUDID UDI-DI 00873263004432 listed 17 Sep 2016 Listed
Recalls
-
- Terminated Z-2350-2018
- Terminated Z-2349-2018
- Terminated Z-2351-2018
- Terminated Z-2352-2018
- Terminated Z-2354-2018
- Terminated Z-2357-2018
- Terminated Z-2358-2018
The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
-
- Terminated Z-0312-2011
- Terminated Z-0314-2011
Fixation thumbscrew on the DBS depth stop adapter may be overtightened and damage the implantable lead.
Timeline
- Recall initiated FDA recalls Z-0312-2011
- Recall initiated FDA recalls Z-0314-2011
- Listed FDA GUDID 00873263004432
- Recall initiated FDA recalls Z-2350-2018
- Recall initiated FDA recalls Z-2349-2018
- Recall initiated FDA recalls Z-2351-2018
- Recall initiated FDA recalls Z-2352-2018
- Recall initiated FDA recalls Z-2354-2018
- Recall initiated FDA recalls Z-2357-2018
- Recall initiated FDA recalls Z-2358-2018
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- STar™ Drive M/E System
- Generic name
- Neurological stereotactic surgery system
- Model / Ref / Catalog
- ST-DS-ME-00443, ST-DS-ME
- Primary DI / UDI-DI
- 00873263004432
- 510(k) Number
- FDA product code
- HAW
- GMDN term
- Neurological stereotactic surgery system
- Device class
- Class II
- Regulation number
- 882.4560
- GUDID distribution status
- In Commercial Distribution
- Sterilisation method
- Moist Heat or Steam Sterilization
- MRI safety
- MR Unsafe
Similar devices
Other devices filed under FDA product code HAW (Neurological stereotaxic Instrument).
- Isg Family Of Viewing Wands by I.S.G. Technologies, Inc.
- STEALTHSTATION S7 / I7 / FUSION NAVIGATION SYSTEMS, AXIEM NAVIGATION SYSTEM by Medtronic Navigation Inc.
- CAROLINE GUIDE/DELIA LOCALIZER by Surgical Solutions, Inc.
- 6003-028-000 by Stryker Leibinger
- GNOMOS STEREOTACTIC SYSTEM W/ NONINVASIVE FIXATION by Nomos Corp.
- VOYAGER 6.0 SOFTWARE OPTION by Philips Medical Systems (Cleveland), Inc.
- 6007-003-100 by Stryker Leibinger
- MEDTRONIC ORTHOPAEDIC TRAUMA APPLICATION by Medtronic Navigation Inc.
- VECTORVISION by Brainlab, Inc.