StealthStation System with Cranial Software
510(k) K153660
StealthStation System with Cranial Software is a medical device (FDA Class II; FDA product code HAW, Neurological Stereotaxic Instrument) made by Medtronic Navigation Inc., cleared under FDA 510(k) K153660 on 14 Sep 2016, active as of 8 Nov 2026. 3 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK153660 | Date14 Sep 2016 | StatusActive |
Official records
- FDA 510(k) database K153660
From the FDA record
- Received by the FDA
- 21 Dec 2015
- FDA decision
- 14 Sep 2016 (Substantially Equivalent)
- FDA review time
- 268 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Ongoing Z-1666-2023
In nonaxial/some axial exams if surgical plan utilized, Target Guidance selected, tip projection utilized, Navigate Projection enabled, software anomaly may occur during tumor resection/shunt placement/deep brain stimulation, where distance to/past target text no longer synchronized with navigation information, inaccurate value displayed; may result in prolonged/additional procedure, tissue injury
-
- Terminated Z-1995-2020
Entry point and lead placement inaccuracies during deep brain stimulation lead implantation procedures may occur when using a specific combination of the firm's Steriotactic System and auto-registration feature with a specific imaging system (also known as a fiducial-less procedure). Minor patient movement may not be initially detected by the user or the software during the auto-registration scan process potentially resulting in inaccuracies and risks for the patient including: inaccurate lead placement, delay of surgery, aborted surgery, or additional intervention (including revision of the lead placement and subsequent imaging).
-
- Terminated Z-0552-2019
An incorrect display of the Biopsy Depth Gauge graphical representation in the Guidance View using Cranial software, which resulted in the biopsy of healthy tissue.
Timeline
- Cleared FDA 510(k) K153660
- Recall initiated FDA recalls Z-0552-2019
- Recall initiated FDA recalls Z-1995-2020
- Recall initiated FDA recalls Z-1666-2023
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- HAW
- FDA applicant
- Medtronic Navigation, Inc.
- Device class
- Class II
- Regulation
- 21 CFR Part 882.4560
- Advisory committee
- Neurology
- FDA decision
- SESE Substantially Equivalent
Devices cleared under K153660
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| StealthStation® Cranial 9735585 | 9735585 | 00643169625532 | — | On the market |
| StealthStation® Cranial 9735585 | 9735585 | 00763000095741 | — | On the market |
| StealthStation® Cranial 9735585 | 9735585 | 00763000253165 | — | On the market |
| StealthStation® Cranial 9735586 | 9735586 | 00643169625570 | — | On the market |
| StealthStation® Cranial 9735586 | 9735586 | 00763000299972 | — | On the market |
| StealthStation® Cranial 9735586 | 9735586 | 00763000517274 | — | On the market |
| StealthStation® Cranial 9735586 | 9735586 | 00763000095758 | — | On the market |
| StealthStation® Cranial 9735587 | 9735587 | 00763000299927 | — | No longer on the market |
| StealthStation® Cranial 9735587 | 9735587 | 00763000517335 | — | On the market |
| StealthStation® Cranial 9735587 | 9735587 | 00643169625631 | — | On the market |
| StealthStation® Cranial 9735587 | 9735587 | 00763000095819 | — | On the market |
Similar devices
Other devices filed under FDA product code HAW (Neurological Stereotaxic Instrument).
- STRYKER ORTHOPAEDICS RFID INSTRUMENTATION by Howmedica Osteonics Corp.
- ClearPoint Array System (version 1.1) by Clearpoint Neuro, Inc.
- ROSA ONE Brain Application by Medtech S.A.S
- STEREOTACTIC ONE MEDICAL APPLICATION SOFTWARE by Stereotactic Image Systems, Inc.
- STEALTHSTATION WITH FLUOROMERGE SOFTWARE by Medtronic Surgical Navigation Technologies
- NAVIGUS UNIBODY FIDUCIAL MARKER SYSTEM, MODEL FM-4000 by Image-Guided Neurologics, Inc.
- NAVIGABLE BRAIN BIOPSY CANNULA SET by IZI MEDICAL PRODUCTS LLC
- MAYFIELD ACCISS II IMAGE GUIDED SURGERY SYSTEM, MODEL 2002-01 by Schaerer Mayfield USA, Inc.
- VECTORVISION SPINE MODULE by Brainlab AG
- 6002-350-000 by Stryker Leibinger