Skip to content

Sterrad 100S Sterilizer

by Advanced Sterilization Products

510(k) K991999

Sterrad 100S Sterilizer is a medical device (FDA Class II; FDA product code MLR, Sterilizer, Chemical) made by Advanced Sterilization Products, cleared under FDA 510(k) K991999 on 15 Jun 2000, active as of 22 Oct 2026. 6 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

6 closed

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK991999 Date15 Jun 2000ClearedStatusActive

Official records

From the FDA record

Received by the FDA
14 Jun 1999
FDA decision
15 Jun 2000 (Substantially Equivalent)
FDA review time
367 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K991999 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

6 recalls · FDA

    • Terminated Z-2337-2012 FDA Class III initiated 19 Jun 2012

    Advanced Sterilization Products (ASP) is recalling (removing) the Sterrad 100S and Sterrad NX System cassettes because the camera used to verify printing on the Sterrad System cassettes was inadvertently disabled. The STERRAD 100S and STERRAD NX System cases were shipped without verification of proper printing.

    • Terminated Z-0340-2009 FDA Unclassified (pre-2012) initiated 17 Oct 2008

    1) Inability of the sterilizer to detect when an injection takes place without hydrogen peroxide being transferred to the vaporizer bowl. This situation can result in a cycle completion with the injection of insufficient hydrogen peroxide sterilant and 2) Inability of the sterilizer to detect an obstruction in the door travel path while the door is closing. This situation can result in a hand b

    • Terminated Z-1628-2008 FDA Unclassified (pre-2012) initiated 25 Feb 2008

    Oil mist filter fails allowing oil mist to be emitted into the vicinity of the sterilizer, which may result in the release of a mist, "haze," or "smoke."

    • Terminated Z-0844-2008 FDA Unclassified (pre-2012) initiated 12 Nov 2007

    Outdated Compatibility Lists: Recalling firm has decided to discontinue dissemination of Compatible Medical Device Reference Lists and instrument assessment activities

    • Terminated Z-0845-2008 FDA Unclassified (pre-2012) initiated 12 Nov 2007

    User Guide Update: based on user reports about contact with residual hydrogen peroxide from instruments, pouches and trays after sterilization and subsequent light colored residue on these devices after sterilization updated user guidance has been issued.

    • Terminated Z-0197-06 FDA Unclassified (pre-2012) initiated 16 Sep 2005

    The STERRAD 100 and the STERRAD 100S vaporizer plates are prone to displacement allowing micro-droplets of H2O2 to contact the load. If the vaporizer plate is dislodged, there is an increased chance for operators to have contact with H2O2 resulting in transient burns during the removal of the load.

Timeline

7 dated events

  1. Cleared FDA 510(k) K991999
  2. Recall initiated FDA recalls Z-0197-06
  3. Recall initiated FDA recalls Z-0844-2008
  4. Recall initiated FDA recalls Z-0845-2008
  5. Recall initiated FDA recalls Z-1628-2008
  6. Recall initiated FDA recalls Z-0340-2009
  7. Recall initiated FDA recalls Z-2337-2012

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
MLR
Device class
Class II
Regulation
21 CFR Part 880.6860
Advisory committee
General Hospital
FDA decision
SESE Substantially Equivalent

Devices cleared under K991999

2 devices · FDA GUDID

Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.

Devices cleared under K991999
UDI-DISize and differences
Sterrad 10101
ADVANCED STERILIZATION PRODUCTS SERVICES INC.
10101 / 1010110705037014460
Single use: no
No longer on the market
STERRAD™ 10113
ADVANCED STERILIZATION PRODUCTS SERVICES INC.
10113 / 1011310705037014583
Single use: yes
No longer on the market

Similar devices

Other devices filed under FDA product code MLR (Sterilizer, Chemical).

Other devices from Advanced Sterilization Products, Inc.

See all devices from Advanced Sterilization Products, Inc. →