Sterrad 10101, a plasma sterilizer (FDA Class II) made by Advanced Sterilization Products Services Inc., was cleared under FDA 510(k) K023290 on 3 Apr 2003 (1 earlier submission) and listed in FDA GUDID on 24 Sep 2016 (UDI-DI 10705037014460); the registry lists it as not in commercial distribution (GUDID) as of 23 Oct 2026. FDA recalls on record: 13.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number10705037014460 | Date24 Sep 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK023290 | Date3 Apr 2003 | StatusActive |
Official records
- AccessGUDID record 10705037014460
- FDA 510(k) database K023290
- FDA 510(k) database K991999
From the FDA record
- Device description
- Sterrad 100S Sterilizer
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA 510(k) K023290 decided 3 Apr 2003 Active
- FDA GUDID UDI-DI 10705037014460 listed 24 Sep 2016 Listed
Recalls
-
- Terminated Z-0842-2015
- Terminated Z-0843-2015
Advanced Sterilization Products is recalling the STERRAD System cassettes because the the leak indicator on the STERRAD System cassettes packaging may not function properly after exposure to hydrogen peroxide.
-
- Terminated Z-0744-2014
- Terminated Z-0745-2014
- Terminated Z-0746-2014
Advanced Sterilization Products (ASP) is recalling the STERRAD NX, STERRAD 50, and STERRAD 100S Sterilization Systems because some healthcare workers detected odors or smells possibly emanating from STERRAD Systems.
-
- Terminated Z-2337-2012
Advanced Sterilization Products (ASP) is recalling (removing) the Sterrad 100S and Sterrad NX System cassettes because the camera used to verify printing on the Sterrad System cassettes was inadvertently disabled. The STERRAD 100S and STERRAD NX System cases were shipped without verification of proper printing.
-
- Terminated Z-1419-2009
ASP has identified a secondary, lower-occurring cause of oil mist within certain STERRAD NX Systems and STERRAD 50 Systems due to potentially defective oil fill plugs.
-
- Terminated Z-0340-2009
1) Inability of the sterilizer to detect when an injection takes place without hydrogen peroxide being transferred to the vaporizer bowl. This situation can result in a cycle completion with the injection of insufficient hydrogen peroxide sterilant and 2) Inability of the sterilizer to detect an obstruction in the door travel path while the door is closing. This situation can result in a hand b
-
- Terminated Z-1628-2008
- Terminated Z-1629-2008
Oil mist filter fails allowing oil mist to be emitted into the vicinity of the sterilizer, which may result in the release of a mist, "haze," or "smoke."
-
- Terminated Z-0844-2008
Outdated Compatibility Lists: Recalling firm has decided to discontinue dissemination of Compatible Medical Device Reference Lists and instrument assessment activities
-
- Terminated Z-0845-2008
User Guide Update: based on user reports about contact with residual hydrogen peroxide from instruments, pouches and trays after sterilization and subsequent light colored residue on these devices after sterilization updated user guidance has been issued.
-
- Terminated Z-0197-06
The STERRAD 100 and the STERRAD 100S vaporizer plates are prone to displacement allowing micro-droplets of H2O2 to contact the load. If the vaporizer plate is dislodged, there is an increased chance for operators to have contact with H2O2 resulting in transient burns during the removal of the load.
Timeline
- Recall initiated FDA recalls Z-1629-2008
- Recall initiated FDA recalls Z-0340-2009
- Recall initiated FDA recalls Z-1419-2009
- Recall initiated FDA recalls Z-2337-2012
- Recall initiated FDA recalls Z-0744-2014
- Recall initiated FDA recalls Z-0745-2014
- Recall initiated FDA recalls Z-0746-2014
- Recall initiated FDA recalls Z-0842-2015
- Recall initiated FDA recalls Z-0843-2015
- Listed FDA GUDID 10705037014460
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Sterrad
- Generic name
- Plasma sterilizer
- Model / Ref / Catalog
- 10101
- Primary DI / UDI-DI
- 10705037014460
- FDA product code
- FLF, MLR
- GMDN term
- Plasma sterilizer
- Device class
- Class II
- Regulation number
- 880.6860
- GUDID distribution status
- Not in Commercial Distribution
Similar devices
Other devices filed under FDA product code FLF (Sterilizer, ethylene-oxide gas).
- ABTOX PLAZLYTE PROCESS MONITOR by Abtox, Inc.
- ARS ENTERPRISES ETHYLENE OXIDE GAS STERILIZER by Ars Enterprises, Inc.
- EOGas 4 Ethylene Oxide Gas Sterilizer by Andersen Sterilizers, Inc.
- GAS STERILIZATION EQUIPMENT by Getinge International, Inc.
- EAGLE 2000 STERILIZER SERIES GAS/AERA by Amsco Co.
- 3M STERI-VAC STERILIZER/AERATOR by 3M COMPANY
- STERRAD 100 STERILIZATION SYSTEM by Johnson & Johnson Medical, Inc.
- MDT/CASTLE 10/90 LOAD STATION by Mdt Biologic Co.
- Anprolene AN75 Ethylene Oxide Gas Sterilizer by Andersen Sterilizers, Inc.
- H & W HOSPITAL ETHYLENE OXIDE STERILIZER CONTROL SYSTEM by H & W Technology, LLC
Other devices from Advanced Sterilization Products, Inc.
See all devices from Advanced Sterilization Products, Inc. →