Symbia 6.5
by Siemens Medical Solutions USA, Inc.
510(k) K162337
Symbia 6.5 is a Class II device made by Siemens Medical Solutions USA, Inc., cleared under FDA 510(k) K162337 on 8 Sep 2016, active as of 22 Oct 2026. 4 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK162337 | Date8 Sep 2016 | StatusActive |
Official records
- FDA 510(k) database K162337
Recalls
-
- Terminated Z-0372-2019
The Evo Excel External-ECG Power-Connection current limit is mis-labeled as 1.0A. It should be labeled 0.55A.,
-
- Terminated Z-0093-2018
- Terminated Z-0094-2018
The Broad Quantification option of the Symbia product software version VB20A may not allow modification of two data input values. The failure occurs when the system with the Broad Quantification option is calibrated with phantoms that do not have a volume of 6500ml and/or when residual dose needs to be entered.
-
- Terminated Z-0047-2018
The torque wrench used to tighten system bolts during installation was found to be out of tolerance
Timeline
- Cleared FDA 510(k) K162337
- Recall initiated FDA recalls Z-0047-2018
- Recall initiated FDA recalls Z-0093-2018
- Recall initiated FDA recalls Z-0094-2018
- Recall initiated FDA recalls Z-0372-2019
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- KPS
- Device class
- Class II
- Regulation
- 21 CFR Part 892.1200
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code KPS (System, Tomography, Computed, Emission).
- NEUROQ 3.6 by Syntermed, Inc.
- PRISM 1500XP NUCLEAR GAMMA CAMERA SYSTEM by Ohio Imaging, Inc.
- MCD-AC by Adac Laboratories
- CONVERGENCE CDR FOR HITACHI SPECTRADIGITAL V250DSP GAMMA CAMERAS by Hitachi Medical Corp. of America
- SKYLIGHT IMAGING SYSTEM by Adac Laboratories
- Precision DL by GE Healthcare
- SIR-Spheres Microspheres Activity Calculator (SMAC) by Sirtex Medical US Holdings, Inc.
- PENN-PET SCANNER by Ugm Medical Systems, Inc.
- VPC-94 ULTRA HIGH ENERGY COLLIMATERS MODEL NUMBERS 480-685-9402 by Ge Medical Systems F.I. Haifa
- EFAI CARDIOSUITE SPECT Myocardial Perfusion Agile Workflows by Ever Fortune.Ai, Co., Ltd.