Symbia 6.0
by Siemens Medical Solutions USA, Inc.
510(k) K142006
Symbia 6.0 is a Class II device made by Siemens Medical Solutions USA, Inc., cleared under FDA 510(k) K142006 on 21 Aug 2014, active as of 22 Oct 2026. 7 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK142006 | Date21 Aug 2014 | StatusActive |
Official records
- FDA 510(k) database K142006
Recalls
-
- Terminated Z-2758-2016
"We have received reports of patients hair being caught in the e.cam and Symbia E patient bed up/down drive. The purpose of this letter is to remind users of the patient positioning instructions for your e.cam or Symbia E system and to inform you of an update that will be performed by your local Siemens service representative."
-
- Terminated Z-2898-2020
E. CAM and Symbia system with foresight detectors performing gated or dynamic acquisition may lose some detector time information. Our risk analysis indicates that the probability of occurrence is remote. SPECT: To detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV. CT: The CT component is intended to produce cross sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial place taken at different angles or spiral planes taken at different angles. SPECT+CT: Perform CT scans and nuclear imaging studies within the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient's anatomy.
-
- Terminated Z-1262-2016
- Terminated Z-1263-2016
We received a customer complaint that Siemens reported via the MedWatch process (1423253-2015-00002). The result of a Hazard Analysis and CAPA investigation, we have determined that for e.cam or Symbia E systems that although the existing user manual is correct, additional supplemental instructions, with the inclusion of visual aids, may further clarify and emphasize the proper collimator change
-
- Terminated Z-0687-2016
To provide operator instruction manuals for the e.cam Dual Signature; Gamma Camera that were not sent in its entirety when the e.cam systems were refurbished and shipped to the customer.
-
- Terminated Z-2437-2015
- Terminated Z-2436-2015
- Terminated Z-2435-2015
Fuses with the improper amperage rating were installed in the Symbia S, Symbia T Series and Symbia lntevo systems. These fuses, in conjunction with a specific failure mode of the Power Controller assembly, could lead to the overheating of the assembly.
-
- Terminated Z-0208-2016
Potential failure mode related to the radial brake, where the detector was able to drift, led to a re-design of the brake redundancy circuit. The redesigned Brake Redundancy Circuit will force even a defective brake to engage in the unpowered state by applying a reverse charge to the brake and ensuring the brake has demagnetized.
Timeline
- Cleared FDA 510(k) K142006
- Recall initiated FDA recalls Z-0208-2016
- Recall initiated FDA recalls Z-2437-2015
- Recall initiated FDA recalls Z-2436-2015
- Recall initiated FDA recalls Z-2435-2015
- Recall initiated FDA recalls Z-0687-2016
- Recall initiated FDA recalls Z-1262-2016
- Recall initiated FDA recalls Z-1263-2016
- Recall initiated FDA recalls Z-2898-2020
- Recall initiated FDA recalls Z-2758-2016
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- KPS
- Device class
- Class II
- Regulation
- 21 CFR Part 892.1200
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Devices cleared under K142006
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| Collimator Asm Smart Zoom 10183368 | 10183368 | 04056869002057 | — | On the market |
| Fan Beam Collimator 04365842 | 4365842.. / 04365842 | 04056869001920 | — | On the market |
| Fan Beam Collimator Asm Symbia 07332062 | 7332062.. / 07332062 | 04056869002019 | — | On the market |
| HI-EN Collimator Asm Symbia 07332054 | 7332054.. / 07332054 | 04056869001982 | — | On the market |
| HI_RES Collimator Asm Symbia 07331973 | 7331973.. / 07331973 | 04056869002026 | — | On the market |
| High Energy Collimator 04355124 | 4355124.. / 04355124 | 04056869001913 | — | On the market |
| LEAP Collimator Asm Symbia 07331882 | 7331882.. / 07331882 | 04056869002002 | — | On the market |
| Low Energy High Res Collimator 04355108 | 4355108.. / 04355108 | 04056869001906 | — | On the market |
| Low Energy Ultra HighRes Collimator 05224931 | 5224931.. / 05224931 | 04056869001951 | — | On the market |
| Low Energy, All Purpose 05224923 | 5224923.. / 05224923 | 04056869001944 | — | On the market |
| Low Pen HR Collimator Symbia E 10764173 | 10764173 | 04056869001975 | — | On the market |
| LPHR COLLIMATOR ASM SYMBIA 10764171 | 10764171 | 04056869002064 | — | On the market |
| Medium Energy Collimator 05232868 | 5232868.. / 05232868 | 04056869001968 | — | On the market |
| Medium Energy Collimator Asm Symbia 07332005 | 7332005.. / 07332005 | 04056869001999 | — | On the market |
| Pin Hole Collimator Asm Symbia 07332039 | 7332039.. / 07332039 | 04056869002033 | — | On the market |
| Pinhole with 4 mm aperture 04380585 | 4380585.. / 04380585 | 04056869001937 | — | On the market |
| UHRES Collimator Asm Symbia 07331957 | 7331957.. / 07331957 | 04056869002040 | — | On the market |
Similar devices
Other devices filed under FDA product code KPS (System, Tomography, Computed, Emission).
- POSITRON EMISSION COMPUTED TOMOGRAPH SET-1400W by Shimadzu Medical Systems
- QPS (QUANTITATIVE PERFUSION SPECT), QGS (QUANTITATIVE GATED SPECT) by Smv America
- QPS/BPGS/MOCO PROCESSING APPLICATIONS FOR ENTEGRA WORKSTATION by Ge Medical Systems F.I. Haifa
- TRIAD NUCLEAR IMAGING SYSTEM by Trionix Research Laboratory, Inc.
- NEUROQ - PET DP by Syntermed, Inc.
- Discovery NM/CT 670 CZT by GE Healthcare
- CEDARS-SINAI QUANTITATIVE THALLIUM-201 ANALYSIS by Nuccardiac Software, Inc.
- WBR XPRESS PET by Ultraspect, Ltd.
- CEQUAL-QUANTITATIVE ANALYSIS by Nuccardiac Software, Inc.
- VANTAGE 2.1 EXSPECT by Adac Laboratories