Skip to content

Symbia 6.0

by Siemens Medical Solutions USA, Inc.

510(k) K142006

Symbia 6.0 is a Class II device made by Siemens Medical Solutions USA, Inc., cleared under FDA 510(k) K142006 on 21 Aug 2014, active as of 22 Oct 2026. 7 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

9 closed

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK142006 Date21 Aug 2014ClearedStatusActive

Official records

Recalls

9 recalls · FDA

    • Terminated Z-2758-2016 FDA Class II initiated 5 Aug 2016 on shared 510(k) K142006

    "We have received reports of patients hair being caught in the e.cam and Symbia E patient bed up/down drive. The purpose of this letter is to remind users of the patient positioning instructions for your e.cam or Symbia E system and to inform you of an update that will be performed by your local Siemens service representative."

    • Terminated Z-2898-2020 FDA Class II initiated 31 Mar 2016

    E. CAM and Symbia system with foresight detectors performing gated or dynamic acquisition may lose some detector time information. Our risk analysis indicates that the probability of occurrence is remote. SPECT: To detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV. CT: The CT component is intended to produce cross sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial place taken at different angles or spiral planes taken at different angles. SPECT+CT: Perform CT scans and nuclear imaging studies within the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient's anatomy.

  • 2 recalls, same reason

    • Terminated Z-1262-2016 FDA Class II initiated 26 Feb 2016
    • Terminated Z-1263-2016 FDA Class II initiated 26 Feb 2016

    We received a customer complaint that Siemens reported via the MedWatch process (1423253-2015-00002). The result of a Hazard Analysis and CAPA investigation, we have determined that for e.cam or Symbia E systems that although the existing user manual is correct, additional supplemental instructions, with the inclusion of visual aids, may further clarify and emphasize the proper collimator change

    • Terminated Z-0687-2016 FDA Class II initiated 31 Dec 2015 on shared 510(k) K142006

    To provide operator instruction manuals for the e.cam Dual Signature; Gamma Camera that were not sent in its entirety when the e.cam systems were refurbished and shipped to the customer.

  • 3 recalls, same reason

    • Terminated Z-2437-2015 FDA Class II initiated 31 Jul 2015
    • Terminated Z-2436-2015 FDA Class II initiated 31 Jul 2015
    • Terminated Z-2435-2015 FDA Class II initiated 31 Jul 2015

    Fuses with the improper amperage rating were installed in the Symbia S, Symbia T Series and Symbia lntevo systems. These fuses, in conjunction with a specific failure mode of the Power Controller assembly, could lead to the overheating of the assembly.

    • Terminated Z-0208-2016 FDA Class II initiated 26 Jun 2015

    Potential failure mode related to the radial brake, where the detector was able to drift, led to a re-design of the brake redundancy circuit. The redesigned Brake Redundancy Circuit will force even a defective brake to engage in the unpowered state by applying a reverse charge to the brake and ensuring the brake has demagnetized.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

10 dated events

  1. Cleared FDA 510(k) K142006
  2. Recall initiated FDA recalls Z-0208-2016
  3. Recall initiated FDA recalls Z-2437-2015
  4. Recall initiated FDA recalls Z-2436-2015
  5. Recall initiated FDA recalls Z-2435-2015
  6. Recall initiated FDA recalls Z-0687-2016
  7. Recall initiated FDA recalls Z-1262-2016
  8. Recall initiated FDA recalls Z-1263-2016
  9. Recall initiated FDA recalls Z-2898-2020
  10. Recall initiated FDA recalls Z-2758-2016

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
KPS
Device class
Class II
Regulation
21 CFR Part 892.1200
Advisory committee
Radiology
FDA decision
SESE Substantially Equivalent

Devices cleared under K142006

17 devices · FDA GUDID

Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.

Devices cleared under K142006
UDI-DISize and differences
Collimator Asm Smart Zoom 10183368
SIEMENS MEDICAL SOLUTIONS USA, INC.
1018336804056869002057 —On the market
Fan Beam Collimator 04365842
SIEMENS MEDICAL SOLUTIONS USA, INC.
4365842.. / 0436584204056869001920 —On the market
Fan Beam Collimator Asm Symbia 07332062
SIEMENS MEDICAL SOLUTIONS USA, INC.
7332062.. / 0733206204056869002019 —On the market
HI-EN Collimator Asm Symbia 07332054
SIEMENS MEDICAL SOLUTIONS USA, INC.
7332054.. / 0733205404056869001982 —On the market
HI_RES Collimator Asm Symbia 07331973
SIEMENS MEDICAL SOLUTIONS USA, INC.
7331973.. / 0733197304056869002026 —On the market
High Energy Collimator 04355124
SIEMENS MEDICAL SOLUTIONS USA, INC.
4355124.. / 0435512404056869001913 —On the market
LEAP Collimator Asm Symbia 07331882
SIEMENS MEDICAL SOLUTIONS USA, INC.
7331882.. / 0733188204056869002002 —On the market
Low Energy High Res Collimator 04355108
SIEMENS MEDICAL SOLUTIONS USA, INC.
4355108.. / 0435510804056869001906 —On the market
Low Energy Ultra HighRes Collimator 05224931
SIEMENS MEDICAL SOLUTIONS USA, INC.
5224931.. / 0522493104056869001951 —On the market
Low Energy, All Purpose 05224923
SIEMENS MEDICAL SOLUTIONS USA, INC.
5224923.. / 0522492304056869001944 —On the market
Low Pen HR Collimator Symbia E 10764173
SIEMENS MEDICAL SOLUTIONS USA, INC.
1076417304056869001975 —On the market
LPHR COLLIMATOR ASM SYMBIA 10764171
SIEMENS MEDICAL SOLUTIONS USA, INC.
1076417104056869002064 —On the market
Medium Energy Collimator 05232868
SIEMENS MEDICAL SOLUTIONS USA, INC.
5232868.. / 0523286804056869001968 —On the market
Medium Energy Collimator Asm Symbia 07332005
SIEMENS MEDICAL SOLUTIONS USA, INC.
7332005.. / 0733200504056869001999 —On the market
Pin Hole Collimator Asm Symbia 07332039
SIEMENS MEDICAL SOLUTIONS USA, INC.
7332039.. / 0733203904056869002033 —On the market
Pinhole with 4 mm aperture 04380585
SIEMENS MEDICAL SOLUTIONS USA, INC.
4380585.. / 0438058504056869001937 —On the market
UHRES Collimator Asm Symbia 07331957
SIEMENS MEDICAL SOLUTIONS USA, INC.
7331957.. / 0733195704056869002040 —On the market

Similar devices

Other devices filed under FDA product code KPS (System, Tomography, Computed, Emission).

Other devices from Siemens Medical Solutions USA, Inc.

See all devices from Siemens Medical Solutions USA, Inc. →