TAS Medical, Inc. Tissue Approximation System T-4000
by TAS Medical, Inc. 2 variants
Listed
Valid until
Not applicable
FDA recalls
1 ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00197644217267 | Date6 Nov 2023 | StatusListed |
Recalls
-
- Ongoing Z-1294-2025
Tissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrent surgery to correct a large herniation.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Trade Name
- TAS Medical, Inc. Tissue Approximation System
- Generic Name
- Nylon suture, non-bioabsorbable, monofilament
- Model / Ref / Catalog
- T-4000
- Primary DI / UDI-DI
- 00197644217267
- 510(k) Number
- FDA Product Code
- GAR
- GMDN Term
- Nylon suture, non-bioabsorbable, monofilament
- Device Class
- Class II
- Regulation Number
- 878.5020
- GUDID distribution status
- Not in Commercial Distribution
- MRI Safety
- MR Safe
- Size
- Length — 37 CentimeterWidth — 2.7 MillimeterHeight — 1.4 Micrometer
Other devices from TAS Medical, Inc.
Properties
Other properties (7)
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No