Triathlon TS Knee System
510(k) K070095
Triathlon TS Knee System is a medical device (FDA Class II; FDA product code JWH) made by Howmedica Osteonics Corp., cleared under FDA 510(k) K070095 on 1 Jun 2007, active as of 22 Oct 2026. 13 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK070095 | Date1 Jun 2007 | StatusActive |
Official records
- FDA 510(k) database K070095
From the FDA record
- Received by the FDA
- 10 Jan 2007
- FDA decision
- 1 Jun 2007 (Substantially Equivalent)
- FDA review time
- 142 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Terminated Z-1635-2015
The peel strength of the inner blister may have been below internal validated requirements. The strength of the seal is directly related to long term shelf life of the sterile barrier.
-
- Terminated Z-1370-2013
Stryker Orthopaedics received a report that a label on the external packaging (box) of a Triathalon Femoral Distal Augment 10mm- Size 5 Left depicted a glyph of an acetabular shell rather than a femoral distal augment. In addition the label was missing a reference to the "10mm" augment thickness which per the label print is required to appear below the size reference.
-
- Terminated Z-0757-2010
- Terminated Z-0758-2010
- Terminated Z-0759-2010
- Terminated Z-0760-2010
- Terminated Z-0761-2010
Triathlon TS femoral components may have been damaged during manufacturing, possibly preventing the assembly of a stem extension.
-
- Terminated Z-1607-2009
- Terminated Z-1608-2009
- Terminated Z-1609-2009
- Terminated Z-1610-2009
Triathlon Offset Adaptors may seize during surgery and the OR staff may be unable to disassemble the instrument using the removal tool.
-
- Terminated Z-0890-2009
Stryker Orthopaedics became aware of a visual defect in certain blister packages of certain devices.
-
- Terminated Z-1231-2007
Stabilizer posts, packaged with the Triathlon TS Plus Tibial inserts have a specification anomaly related to length, resulting in a potential for the post to slightly protrude above the top surface of the plastic post.
Timeline
- Recall initiated FDA recalls Z-1608-2009
- Recall initiated FDA recalls Z-1609-2009
- Recall initiated FDA recalls Z-1610-2009
- Recall initiated FDA recalls Z-0757-2010
- Recall initiated FDA recalls Z-0758-2010
- Recall initiated FDA recalls Z-0759-2010
- Recall initiated FDA recalls Z-0760-2010
- Recall initiated FDA recalls Z-0761-2010
- Recall initiated FDA recalls Z-1370-2013
- Recall initiated FDA recalls Z-1635-2015
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- JWH
- Device class
- Class II
- Regulation
- 21 CFR Part 888.3560
- Advisory committee
- Orthopedic
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code JWH (Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer).
- APEX Knee System EP-22313 by OMNIlife science, Inc.
- APEX Revision Knee KP-4215L by OMNIlife science, Inc.
- FEMORAL AND TIBIAL PLUGS OF INTERAX TOTAL KNEE SYSTEM by Howmedica Corp.
- CONTINUUM KNEE SYSTEM (CKS) by Techmedica, Inc.
- GMK Sphere Total Knee System 02.12.0510CRL by Medacta International SA
- PHYSICA LMC 6537.54.916 by Limacorporate
- PHYSICA KR 6531.54.916 by Limacorporate
- GMK Sphere Total Knee System 02.12.E0210FL by Medacta International SA
- Unity Total Knee System 112.014.84 by CORIN LTD