Trilogy Evo

by Respironics, Inc.

510(k) K181166

Trilogy Evo is a medical device (FDA Class II; FDA product code CBK, Ventilator, Continuous, Facility Use) made by Respironics, Inc., cleared under FDA 510(k) K181166 on 18 Jul 2019, active as of 22 Oct 2026. 33 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

23 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK181166 Date18 Jul 2019ClearedStatusActive

Official records

From the FDA record

Received by the FDA
2 May 2018
FDA decision
18 Jul 2019 (Substantially Equivalent)
FDA review time
442 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K181166 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

36 recalls · 23 ongoing · FDA

    • Ongoing Z-1614-2026 FDA Class I initiated 2 Mar 2026

    Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient.

  • 3 recalls, same reason

    • Ongoing Z-1646-2026 FDA Class I initiated 2 Mar 2026
    • Ongoing Z-1647-2026 FDA Class I initiated 2 Mar 2026
    • Ongoing Z-1648-2026 FDA Class I initiated 2 Mar 2026

    In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.

  • 3 recalls, same reason

    • Ongoing Z-0290-2025 FDA Class I initiated 7 Oct 2024
    • Ongoing Z-0291-2025 FDA Class I initiated 7 Oct 2024
    • Ongoing Z-0293-2025 FDA Class I initiated 7 Oct 2024

    Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.

  • 5 recalls, same reason

    • Ongoing Z-1499-2024 FDA Class I initiated 19 Mar 2024
    • Ongoing Z-1500-2024 FDA Class I initiated 19 Mar 2024
    • Ongoing Z-1501-2024 FDA Class I initiated 19 Mar 2024
    • Ongoing Z-1502-2024 FDA Class I initiated 19 Mar 2024
    • Ongoing Z-1503-2024 FDA Class I initiated 19 Mar 2024

    Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.

    • Ongoing Z-0638-2024 FDA Class II initiated 27 Nov 2023

    The following was missing from the Contraindications Statement: The AVAPS-AE therapy mode is contraindicated for invasive use and patients less than 10 kg. The AVAPS feature is contraindicated for patients less than 10 kg. Potential harms if devices are used on contraindicated patients include barotrauma, hypoventilation/hypercapnia, and rebreathing of excessive CO2.

    • Ongoing Z-2106-2023 FDA Class II initiated 25 May 2023

    There are errors present in the Korean Instructions For Use where the word "invasive," was incorrectly translated to "noninvasive".

  • 3 recalls, same reason

    • Ongoing Z-1649-2023 FDA Class I initiated 1 May 2023
    • Ongoing Z-1650-2023 FDA Class I initiated 1 May 2023
    • Ongoing Z-1651-2023 FDA Class I initiated 1 May 2023

    Environmental debris may accumulate on the internal machine flow sensor causing partial occlusion which may impact accurate delivery of pressure, volume, or flow. A malfunction caused by this issue can result in patient harm up to hypoxemia if not addressed by the care provider.

  • 2 recalls, same reason

    • Ongoing Z-1287-2023 FDA Class I initiated 1 Mar 2023
    • Ongoing Z-1288-2023 FDA Class I initiated 1 Mar 2023

    The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value higher than the device is actually delivering. This may lead to under delivery of oxygen.

    • Ongoing Z-0494-2022 FDA Class I initiated 22 Dec 2021

    Philips has identified a specific lot of non-conforming material manufactured by one of its suppliers. A Philips supplier incorrectly used polyester-based polyurethane (PE-PUR) sound abatement foam, a non-conforming material in the devices and repair kits. PE-PUR foam may degrade into particles which may enter the device s air pathway and be ingested or inhaled by the user. The foam degradation may be exacerbated by use of unapproved cleaning methods such as unapproved ozone cleaners, and the PE-PUR foam may off-gas certain chemicals during initial use of the device.

  • 3 recalls, same reason

    • Ongoing Z-2287-2021 FDA Class II initiated 26 Jul 2021 on shared 510(k) K181166
    • Ongoing Z-2288-2021 FDA Class II initiated 26 Jul 2021 on shared 510(k) K181166
    • Ongoing Z-2289-2021 FDA Class II initiated 26 Jul 2021 on shared 510(k) K181166

    Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).

  • 13 recalls, same reason

    • Terminated Z-0942-2020 FDA Class I initiated 19 Nov 2019
    • Terminated Z-0943-2020 FDA Class I initiated 19 Nov 2019
    • Terminated Z-0944-2020 FDA Class I initiated 19 Nov 2019
    • Terminated Z-0945-2020 FDA Class I initiated 19 Nov 2019
    • Terminated Z-0946-2020 FDA Class I initiated 19 Nov 2019
    • Terminated Z-0947-2020 FDA Class I initiated 19 Nov 2019
    • Terminated Z-0948-2020 FDA Class I initiated 19 Nov 2019
    • Terminated Z-0949-2020 FDA Class I initiated 19 Nov 2019
    • Terminated Z-0950-2020 FDA Class I initiated 19 Nov 2019
    • Terminated Z-0951-2020 FDA Class I initiated 19 Nov 2019
    • Terminated Z-0952-2020 FDA Class I initiated 19 Nov 2019
    • Terminated Z-0953-2020 FDA Class I initiated 19 Nov 2019
    • Terminated Z-0954-2020 FDA Class I initiated 19 Nov 2019

    Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing requirements TSRS1298 and PRD439.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

37 dated events

  1. Recall initiated FDA recalls Z-1501-2024
  2. Recall initiated FDA recalls Z-1502-2024
  3. Recall initiated FDA recalls Z-1503-2024
  4. Recall initiated FDA recalls Z-0290-2025
  5. Recall initiated FDA recalls Z-0291-2025
  6. Recall initiated FDA recalls Z-0293-2025
  7. Recall initiated FDA recalls Z-1614-2026
  8. Recall initiated FDA recalls Z-1646-2026
  9. Recall initiated FDA recalls Z-1647-2026
  10. Recall initiated FDA recalls Z-1648-2026

27 earlier events are in the sections above.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
CBK
Device class
Class II
Regulation
21 CFR Part 868.5895
Advisory committee
Anesthesiology
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code CBK (Ventilator, Continuous, Facility Use).

Other devices from Respironics, Inc.

See all devices from Respironics, Inc. →