Trilogy Evo
510(k) K181166
Trilogy Evo is a medical device (FDA Class II; FDA product code CBK, Ventilator, Continuous, Facility Use) made by Respironics, Inc., cleared under FDA 510(k) K181166 on 18 Jul 2019, active as of 22 Oct 2026. 33 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
23 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK181166 | Date18 Jul 2019 | StatusActive |
Official records
- FDA 510(k) database K181166
From the FDA record
- Received by the FDA
- 2 May 2018
- FDA decision
- 18 Jul 2019 (Substantially Equivalent)
- FDA review time
- 442 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Ongoing Z-1614-2026
Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient.
-
- Ongoing Z-1646-2026
- Ongoing Z-1647-2026
- Ongoing Z-1648-2026
In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.
-
- Ongoing Z-0290-2025
- Ongoing Z-0291-2025
- Ongoing Z-0293-2025
Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.
-
- Ongoing Z-1499-2024
- Ongoing Z-1500-2024
- Ongoing Z-1501-2024
- Ongoing Z-1502-2024
- Ongoing Z-1503-2024
Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.
-
- Ongoing Z-0638-2024
The following was missing from the Contraindications Statement: The AVAPS-AE therapy mode is contraindicated for invasive use and patients less than 10 kg. The AVAPS feature is contraindicated for patients less than 10 kg. Potential harms if devices are used on contraindicated patients include barotrauma, hypoventilation/hypercapnia, and rebreathing of excessive CO2.
-
- Ongoing Z-2106-2023
There are errors present in the Korean Instructions For Use where the word "invasive," was incorrectly translated to "noninvasive".
-
- Ongoing Z-1649-2023
- Ongoing Z-1650-2023
- Ongoing Z-1651-2023
Environmental debris may accumulate on the internal machine flow sensor causing partial occlusion which may impact accurate delivery of pressure, volume, or flow. A malfunction caused by this issue can result in patient harm up to hypoxemia if not addressed by the care provider.
-
- Ongoing Z-1287-2023
- Ongoing Z-1288-2023
The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value higher than the device is actually delivering. This may lead to under delivery of oxygen.
-
- Ongoing Z-0494-2022
Philips has identified a specific lot of non-conforming material manufactured by one of its suppliers. A Philips supplier incorrectly used polyester-based polyurethane (PE-PUR) sound abatement foam, a non-conforming material in the devices and repair kits. PE-PUR foam may degrade into particles which may enter the device s air pathway and be ingested or inhaled by the user. The foam degradation may be exacerbated by use of unapproved cleaning methods such as unapproved ozone cleaners, and the PE-PUR foam may off-gas certain chemicals during initial use of the device.
-
- Ongoing Z-2287-2021
- Ongoing Z-2288-2021
- Ongoing Z-2289-2021
Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).
-
- Terminated Z-0942-2020
- Terminated Z-0943-2020
- Terminated Z-0944-2020
- Terminated Z-0945-2020
- Terminated Z-0946-2020
- Terminated Z-0947-2020
- Terminated Z-0948-2020
- Terminated Z-0949-2020
- Terminated Z-0950-2020
- Terminated Z-0951-2020
- Terminated Z-0952-2020
- Terminated Z-0953-2020
- Terminated Z-0954-2020
Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing requirements TSRS1298 and PRD439.
Timeline
- Recall initiated FDA recalls Z-1501-2024
- Recall initiated FDA recalls Z-1502-2024
- Recall initiated FDA recalls Z-1503-2024
- Recall initiated FDA recalls Z-0290-2025
- Recall initiated FDA recalls Z-0291-2025
- Recall initiated FDA recalls Z-0293-2025
- Recall initiated FDA recalls Z-1614-2026
- Recall initiated FDA recalls Z-1646-2026
- Recall initiated FDA recalls Z-1647-2026
- Recall initiated FDA recalls Z-1648-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- CBK
- Device class
- Class II
- Regulation
- 21 CFR Part 868.5895
- Advisory committee
- Anesthesiology
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code CBK (Ventilator, Continuous, Facility Use).
- MICROPROCESSOR 7200A VENTILATOR by Puritan Bennett Corp.
- Ventoux VC2 VX-08-1-0N1 by FLIGHT MEDICAL INNOVATIONS LTD
- SPACELABS ARKON ANESTHESIA WORKSTATION by Spacelabs Healthcare Inc.
- MVP-1500 CONTINUOUS VENTILATOR by Medimex, Inc.
- SECHRIST MODEL 990 JET VENTILATOR by Sechrist Industries Inc
- SERVO-I VENTILATOR SYSTEM by Maquet Critical Care AB
- Astral 100/150 by Resmed Corp
- V200 INTELLI-TRAK OPTION by Respironics California, Inc.
- SEHRIST VOLUME VENTILATOR, MODEL 2200B by Sechrist Industries Inc
- ENGSTROM CARESTATION, ENGSTROM PRO by Datex-Ohmeda Inc.