Trilogy R1119867
Trilogy R1119867 is a portable electric ventilator (FDA Class I) made by Respironics Inc., cleared under FDA 510(k) K083526 on 13 Mar 2009 and listed in FDA GUDID since 26 Feb 2020 (UDI-DI 00606959038851), active as of 23 Oct 2026. 8 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
3 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00606959038851 | Date26 Feb 2020 | StatusListed |
| United States | FDA 510(k)SESE | NumberK083526 | Date13 Mar 2009 | StatusActive |
Official records
- AccessGUDID record 00606959038851
- FDA 510(k) database K083526
From the FDA record
- Device description
- Trilogy EC Ventilator, Rental, USA
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA 510(k) K083526 decided 13 Mar 2009 Active
- FDA GUDID UDI-DI 00606959038851 listed 26 Feb 2020 Listed
Recalls
-
- Ongoing Z-0882-2023
- Ongoing Z-0884-2023
The replacement silicone sound abatement foam installed into some Trilogy 100 and Trilogy 200 devices may separate from the plastic backing to which it is attached. If this were to happen, the foam could potentially block air inlet, which could result in a reduction in delivered therapy volume or pressure and could cause the device to alarm.
-
- Ongoing Z-1956-2021
The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the devices air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.
-
- Terminated Z-1065-2016
Software Issue
-
- Terminated Z-1143-2014
Internal testing revealed a potentially defective component on the power management board that could affect the function of certain Trilogy Ventilators resulting in loss of therapy to patients.
-
- Terminated Z-2024-2012
The specified Trilogy units are being recalled because of the potential for a faulty component in the power supply.
-
- Terminated Z-0296-2012
Trilogy 100, 200 and 202 units are being recalled due to low impeller press-force used during the manufacturing process. If the press force were low enough, theoretically the impeller could move up the shaft and come into contact with the inside of the blower housing/impeller enclosure. This may impact therapy.
-
- Terminated Z-1896-2010
It is possible that the battery could electrically short, resulting in a potential for fire or a thermal event.
Timeline
- Cleared FDA 510(k) K083526
- Recall initiated FDA recalls Z-1896-2010
- Recall initiated FDA recalls Z-0296-2012
- Recall initiated FDA recalls Z-2024-2012
- Recall initiated FDA recalls Z-1143-2014
- Recall initiated FDA recalls Z-1065-2016
- Listed FDA GUDID 00606959038851
- Recall initiated FDA recalls Z-1956-2021
- Recall initiated FDA recalls Z-0882-2023
- Recall initiated FDA recalls Z-0884-2023
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Trilogy
- Generic name
- Portable electric ventilator
- Model / Ref / Catalog
- R1119867
- Primary DI / UDI-DI
- 00606959038851
- 510(k) Number
- FDA product code
- OUG, DQA, CBK
- GMDN term
- Portable electric ventilator
- Device class
- Class I
- Regulation number
- 880.6310, 870.2700, 868.5895
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code OUG (Medical device data system).
- THD Procto Software System 760004 by THD SPA
- APIWEB™ 1.4 by BIOMERIEUX SA
- 4605000000 4605-000-000 by Stryker Corp.
- Standard Imaging QA Pilot 91348 by STANDARD IMAGING INC.
- Echolaser Smart Interface (ESI) EI KT0 0005 521 by ELESTA SPA
- STIMPOD X-LINK XT-45300 by XAVANT TECHNOLOGY (PTY) LTD
- EDIS 6914 by Carl Zeiss Meditec AG
- seca 452 4520060009 by seca GmbH & Co. KG
- T:CONNECT DATA MANAGEMENT SYSTEM by Tandem Diabetes Care, Inc.
- WD300 by Karl Storz SE & CO. KG