Trulink 90496
by Spacelabs Healthcare (Washington), Inc
Trulink 90496 is a Class II device made by Spacelabs Healthcare (Washington), Inc, cleared under FDA 510(k) K103142 on 3 Mar 2011 and listed in FDA GUDID since 11 Oct 2016 (UDI-DI 10841522110399), active (FDA 510(k)), not in commercial distribution (GUDID) as of 23 Oct 2026. 9 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
8 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number10841522110399 | Date11 Oct 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK103142 | Date3 Mar 2011 | StatusActive |
Official records
- AccessGUDID record 10841522110399
- FDA 510(k) database K103142
Across registries
- FDA 510(k) K103142 decided 3 Mar 2011 Active
- FDA GUDID UDI-DI 10841522110399 listed 11 Oct 2016 Listed
Recalls
-
- Ongoing Z-2245-2026
- Ongoing Z-2246-2026
- Ongoing Z-2247-2026
- Ongoing Z-2244-2026
- Ongoing Z-2243-2026
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
-
- Ongoing Z-2052-2026
Due to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester).
-
- Ongoing Z-1793-2026
Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.
-
- Ongoing Z-1503-2023
Command modules will display the following when parameter processing is suspended, depending on software version: 1) Older versions will have waveforms present with no alarms, 2) Newer versions have waveforms absent with no alarms. This may cause clinicians to miss a potentially catastrophic event.
-
- Terminated Z-2343-2016
The firm has received multiple reports of the Non-Invasive Blood Pressure (NIBP) parameter becoming non-functional with an associated message (No Reading). In addition, there is one of the following three error messages: Inflate Error, HW Error, or No Data.
-
- Terminated Z-1577-2015
Modules resetting. This failure mode happens when the Spacelabs Command Module is configured with the Masimo SpO2 option (-M), the Spacelabs Respiration option (-R) and the Masimo SpO2 PCBA, PN: 010-1136-02. With these two options and the Masimo SpO2 PCBA present, the module may experience random resets.
-
- Terminated Z-1542-2015
Ultraview SL Command Modules which were upgraded with the Masimo SpO2 Option, Model 91496-M was affected. Customer reported that during the time that the module is resetting (~10 seconds), monitoring of all parameters will be suspended. Upon completion of this sequence, alarms will be reset to their default values.
-
- Terminated Z-1705-2014
A software anomaly causes temperature alarm limits to change to limits which are not physiologically possible and result in nuisance false positive alarms.
-
- Terminated Z-0690-2012
NIBP pump inside the Ultraview SL Command Module, Model 91496, became loose. The pump is suspended by the air hoses and electrical connections, but there can be enough movement to damage internal components and cause the module to fail.
-
- Terminated Z-0776-2012
One Spacelabs Medical Ultraview SL Command Module, Model 91496, was removed from customer due to software error.
Timeline
- Recall initiated FDA recalls Z-2343-2016
- Listed FDA GUDID 10841522110399
- Recall initiated FDA recalls Z-1503-2023
- Recall initiated FDA recalls Z-1793-2026
- Recall initiated FDA recalls Z-2052-2026
- Recall initiated FDA recalls Z-2245-2026
- Recall initiated FDA recalls Z-2246-2026
- Recall initiated FDA recalls Z-2247-2026
- Recall initiated FDA recalls Z-2244-2026
- Recall initiated FDA recalls Z-2243-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Trulink
- Generic name
- Electrocardiographic lead set, reusable
- Model / Ref / Catalog
- 90496
- Primary DI / UDI-DI
- 10841522110399
- 510(k) Number
- FDA product code
- DSI
- GMDN term
- Electrocardiographic lead set, reusable
- Device class
- Class II
- Regulation number
- 870.1025
- GUDID distribution status
- Not in Commercial Distribution
Similar devices
Other devices filed under FDA product code DSI (DETECTOR AND ALARM, ARRHYTHMIA).
- Q-TEL TELEMETRY SYSTEM V. 6.0 (ST) by Quinton, Inc.
- ZYMED TELEMETRY SYSTEM: MODEL EASIVIEW by Agilent Technologies Inc.
- ECG Mini System Continuous ECG Monitor and Arrhythmia Detector by Lifewatch Technologies, Ltd.
- QUEST EXERCISE STRESS SYSTEM, MODEL CONFIGURATION 14 by Spacelabs Burdick, Inc.
- OXFORD INTENSIVE CARE PATIENT MON. SYS by Oxford Medilog, Inc.
- QUINTON Q2000 ETT MONITORING SYSTEM by Quinton, Inc.
- MULTI-PARAMETER MODULE (M3000A)/DISPLAY UNIT (M3046A) by Hewlett-Packard Co.
- HP M2600A VIRIDIA TELEMETRY SYSTEM by Hewlett-Packard Co.
- TYPE 9210 S/ECG by S & W Medico Teknik A/S
- INTELLIVUE PATIENT MONITOR, MODELS MP2, MP5, MP20, MP30, MP40, MP50, MO60, MP7-, MP80, AND MP90 by Philips Medizinsysteme Boeblingen GmbH, Cardiac AN
Other devices from Spacelabs Healthcare (Washington), Inc
See all devices from Spacelabs Healthcare (Washington), Inc →