V60 Ventilator, Model V8000
by Respironics California, Inc.
510(k) K082660
V60 Ventilator, Model V8000 is a medical device (FDA Class II; FDA product code MNT, Continuous, Minimal Ventilatory Support, Facility Use Ventilator) made by Respironics California, Inc., cleared under FDA 510(k) K082660 on 9 Jan 2009, active as of 22 Oct 2026. 44 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
22 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK082660 | Date9 Jan 2009 | StatusActive |
Official records
- FDA 510(k) database K082660
From the FDA record
- Received by the FDA
- 12 Sep 2008
- FDA decision
- 9 Jan 2009 (Substantially Equivalent)
- FDA review time
- 119 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Ongoing Z-1137-2021
Due to unqualified "design verification" batteries being shipped to customers.
-
- Ongoing Z-0916-2021
- Ongoing Z-0908-2021
- Ongoing Z-0911-2021
- Ongoing Z-0910-2021
- Ongoing Z-0906-2021
- Ongoing Z-0907-2021
- Ongoing Z-0917-2021
- Ongoing Z-0909-2021
- Ongoing Z-0903-2021
- Ongoing Z-0904-2021
- Ongoing Z-0912-2021
- Ongoing Z-0914-2021
- Ongoing Z-0915-2021
- Ongoing Z-0905-2021
- Ongoing Z-0913-2021
- Ongoing Z-0902-2021
Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting entries and adjustments.
-
- Ongoing Z-1622-2020
Solder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blower to lose power, spool down, and trigger visual and audible High Priority Check Vent alarm indicating a need for alternative ventilation, or 2) Intermittent connection may disrupt expected operation and trigger unexpected shutdown without alarm. Failures may lead to hypercarbia/hypoxemia
-
- Ongoing Z-1155-2020
- Ongoing Z-1153-2020
- Ongoing Z-1154-2020
Ventilator touchscreen may become frozen and fail to respond to touch commands. If a patient requires a change in therapy, there is no warning associated with an unresponsive touchcreen, and therefore the clinician would have no prior indication of a touchscreen failure and would be unable to change ventilator settings. The ventilator will continue to function at the predetermined and accepted settings required to support the patient. A delay in therapy adjustments may result in hypercarbia and/or hypoxemia.
-
- Ongoing Z-1152-2020
Ventilator touchscreen may become frozen and fail to respond to touch commands. If a patient requires a change in therapy, there is no warning associated with an unresponsive touchscreen, and therefore the clinician would have no prior indication of a touchscreen failure and would be unable to change ventilator settings. The ventilator will continue to function at the predetermined and accepted settings required to support the patient. A delay in therapy adjustments may result in hypercarbia and/or hypoxemia.
-
- Terminated Z-0885-2021
- Terminated Z-0899-2021
- Terminated Z-0889-2021
- Terminated Z-0901-2021
- Terminated Z-0898-2021
- Terminated Z-0888-2021
- Terminated Z-0895-2021
- Terminated Z-0886-2021
- Terminated Z-0900-2021
- Terminated Z-0890-2021
- Terminated Z-0891-2021
- Terminated Z-0887-2021
- Terminated Z-0894-2021
- Terminated Z-0892-2021
- Terminated Z-0897-2021
- Terminated Z-0893-2021
- Terminated Z-0896-2021
Ventilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to over discharge, or 2) if the ventilator is stored for an extended period of time without being plugged into AC power.
-
- Terminated Z-2061-2017
Over time, low-frequency vibrations can cause the pins within the female connectors on the Motor Controller to Data Acquisition Board ribbon cable to become partially displaced causing momentary high resistance that prevents or restricts data travelling through the cable during the high resistance event.
-
- Terminated Z-1537-2013
Respironics California, Inc. has initiated a recall on the V60 ventilator because of an issue with the software on the V60 Power Management Board Assembly. If a component fails on the Power Management Board Assembly, it may cause ventilator support to be lost with potentially no audible alarm from the ventilator.
-
- Terminated Z-2260-2012
Respironics is recalling the V60 ventilator because certain blower motor assemblies in the V60 ventilators may not meet specifications. Specifically, the impeller may not be properly seated on the blower motor shaft. This could subsequently cause the blower to cease functioning which would result in the failure of the V60 ventilator to deliver therapy to the patient due to the loss of ventilatio
-
- Terminated Z-0457-2011
Respironics California Inc has received customer complaints of an odor or burning smell coming from the V60 Ventilator, and identified a component on the Power Management PCB that exhibited thermal damage.
-
- Terminated Z-0176-2010
The recall was initiated to correct an issue with a component on the V60 Ventilator Power Management (PM) Printed Circuit Board (PCB). A component on the PCB has the potential to break off or become damaged if the PCB is removed from the ventilator. If this occurs, the ventilator may not properly switch between back up battery power and AC power. This can result in power loss during ventilation fo
Timeline
- Recall initiated FDA recalls Z-0909-2021
- Recall initiated FDA recalls Z-0903-2021
- Recall initiated FDA recalls Z-0904-2021
- Recall initiated FDA recalls Z-0912-2021
- Recall initiated FDA recalls Z-0914-2021
- Recall initiated FDA recalls Z-0915-2021
- Recall initiated FDA recalls Z-0905-2021
- Recall initiated FDA recalls Z-0913-2021
- Recall initiated FDA recalls Z-0902-2021
- Recall initiated FDA recalls Z-1137-2021
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- MNT
- Device class
- Class II
- Regulation
- 21 CFR Part 868.5895
- Advisory committee
- Anesthesiology
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code MNT (Continuous, Minimal Ventilatory Support, Facility Use Ventilator).
- Fisher & Paykel Healthcare 900PT573US by FISHER & PAYKEL HEALTHCARE, INC.
- BIPAP VISION VENTILATORY SUPPORT by Respironics, Inc.
- BiPAP A40 INX3100T19 by Respironics Inc.
- ACUTE CARE FACE MASK, MODEL RT040 by FISHER & PAYKEL HEALTHCARE, INC.
- OmniLab Direct, Web Software - 2 1113127 by Respironics Inc.
- STELLAR 150 by Resmed Germany, Inc.
- SULLIVAN VPAP II ST VENTILATORY SUPPORT SYSTEM by Resmed Corp
- STICK-ON DISPOSABLE NASAL MASK, MODEL 312174, 1001964 by Respironics, Inc.
- BREAS VIVO BILEVEL SYSTEM, MODEL VIVO 40 by Breas Medical AB
- SV70 Ventilator by Shenzhen Mindray Bio-Medical Electronics Co., Ltd.