VarioSurg SG1A
VarioSurg SG1A is a dental ultrasonic surgical system handpiece tip (FDA Class II) made by Nakanishi Inc., cleared under FDA 510(k) K073678 on 10 Jul 2008 and listed in FDA GUDID since 16 Sep 2016 (UDI-DI 04560264563636), active as of 23 Oct 2026. One FDA recall is linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number04560264563636 | Date16 Sep 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK073678 | Date10 Jul 2008 | StatusActive |
Official records
- AccessGUDID record 04560264563636
- FDA 510(k) database K073678
From the FDA record
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile; sterilise before use (Moist Heat or Steam Sterilization)
Across registries
- FDA 510(k) K073678 decided 10 Jul 2008 Active
- FDA GUDID UDI-DI 04560264563636 listed 16 Sep 2016 Listed
Recalls
-
- Terminated Z-2673-2016
Through an error, the labeling failed to include the symbol or other text indicating that the diamond coated tips were single use only. A contributing factor to this error is that in other markets such as Japan, the diamond coated tips are multiple use and are not limited to single use only. The tips are stainless steel with a coating of fine diamond powder. The tips are sold non-sterile and are to be sterilized prior to first use. The tips were cleared as single use only. This problem was discovered during an ongoing, proactive, quality system improvement plan that was initiated by Nakanishi in October 2014. Because the correct instructions for sterilization are not available, it is possible that the devices might not be sterile prior to use. The absence of the single use statement introduces a risk of cross-contamination if the tips are reprocessed and reused.
Timeline
- Cleared FDA 510(k) K073678
- Recall initiated FDA recalls Z-2673-2016
- Listed FDA GUDID 04560264563636
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- VarioSurg
- Generic name
- Dental ultrasonic surgical system handpiece tip
- Model / reference
- SG1A
- Primary DI / UDI-DI
- 04560264563636
- 510(k) Number
- FDA product code
- ELC, DZI
- GMDN term
- Dental ultrasonic surgical system handpiece tip
- Device class
- Class II
- Regulation number
- 872.4850, 872.4120
- GUDID distribution status
- In Commercial Distribution
- Sterilisation method
- Moist Heat or Steam Sterilization
Similar devices
Other devices filed under FDA product code ELC (Scaler, ultrasonic).
- BIOSONIC SUVI PIEZEOELECTRIC SCALER AND POLISHER by Coltene/Whaledent Inc.
- SAPPHIRE PLUS SCALER MOD. 300114/PLUGGER MOD. 300115/CUTTING MOD. 300116/SPREADER MOD. 300117 TIPS by San Diego Swiss Machining, Inc.
- SCORPION ULTRASONIC SCALING INSERTS by SCORPION
- BONART ART-P3II & P4 PIEZO ELECTRIC ULTRASONIC SCALER UNITS WITH ACCESSORIES by BONART CO., LTD.
- PIEZOTOME by Satelec
- SONICFLEX 2003 AND SONICFLEX LUX 2003 L WITH INSTRUMENT TIPS by Sybron Dental Specialties, Inc.
- ENDOCENTER by Satelec - Acteon Group
- MODEL G-61 ULTRASONIC DENTAL UNIT by Cavitron Corp.
- GENTEEL by Luben Dental Co., Inc.
- CAVITRON RF ULTRASONIC SCALER SYSTEM WITH STERIMATE HANDPIECE, MODEL G-131 by Dentsply Intl.