Verigene C. Difficile Nucleic Acid Test (Cdf)
by Nanosphere, Inc. 510(k)
Active
510(k) K123197
Regulatory status
Recall ongoing. 2 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK123197 | Date5 Dec 2012 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-2024-2023
It is possible that in rare instances the VERIGENE CDF Nucleic Acid Test Kit (Part No. 20-005-022) and the VERIGENE Enteric Pathogens Nucleic Acid Test Kit (Part No. 20-005-023) could provide a false-negative result due to hydrophobic characteristics in specific FLOQSwab lots.
-
- Ongoing Z-2010-2021
Potential for a false-negative result.
-
- Terminated Z-2168-2014
Nanosphere has received report of an increased rate of "No Call - INT CTL 2" results for the Internal Control 2 (INT CTL 2) when negative samples are tested with certain lots of CDF Amplification Trays (20-011-022), which are contained in the CDF Amplification Reagent Kit (20-012-022); specifically for lots 061913022D and 071813022E.
-
- Terminated Z-2682-2014
Nanosphere has received several reports relating to an increased rate of Processing Errors during procedures performed on the Processor SP caused by Tips in Tip Holder Assemblies for the following products and lots of Extraction Trays.
Specification
- 510(k) Number
- FDA Product Code
- OZN
- Device Class
- Class II
- Regulation
- 21 CFR Part 866.3130
- Advisory Committee
- Microbiology
- FDA Decision
- SESE Substantially Equivalent