Verigene Gram Negative Blood Culture Nucleic Acid Test (Bc-Gn)

by Nanosphere, Inc.

510(k) K132843

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

2 ongoing

Recall on shared 510(k) K132843; other models may be affected

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK132843 Date8 Jan 2014ClearedStatusActive

Recalls

8 recalls · 2 ongoing · FDA

    • Ongoing Z-1109-2022 FDA Class II initiated 15 Feb 2022 on shared 510(k) K132843

    Higher than normal "No Call Internal Control failures" due to a missing reagent within a small subset of the extraction tray lot.

    • Ongoing Z-2008-2021 FDA Class II initiated 30 Oct 2020 on shared 510(k) K132843

    Potential for a false-negative result.

    • Completed Z-1903-2025 FDA Class II initiated 16 Apr 2025

    It was determined that the consumable issue may result in No Call - INT CTL 1 and/or False Negative results on a small portion of the lot of test cartridges.

    • Terminated Z-2776-2020 FDA Class II initiated 15 Jun 2020

    An instrument failure mode that may result in a Blood Culture Gram Positive (BC-GP) or Gram Negative (BC-GN) false negative call.

    • Terminated Z-2236-2014 FDA Class II initiated 9 Jul 2014

    Nanosphere Inc. has recently determined through four customer complaints that a small number of Verigene Gram-Negative Blood Culture Nucleic Acid Test (BC-GN) Extraction Trays in Lot 042914021B have low levels or are missing Magnetic Microparticles (MMPs) required for DNA Extraction.

    • Terminated Z-2170-2014 FDA Class II initiated 17 Apr 2014

    Nanosphere has determined that a portion of BC-GN Test Cartridge Lot number 040214021A has an elevated risk for No Call results and False Negative results. The BC-GN Test Cartridge (20-006-021) is distributed as part of the Verigene BC-GN Test Kit (20-005-021), which is labeled as "IVD". Therefore, Nanosphere has initiated a voluntary recall to retrieve the following BC-GN Test Cartridges: Test Ca

    • Terminated Z-2681-2014 FDA Class II initiated 10 Oct 2013

    Nanosphere has received several reports relating to an increased rate of Processing Errors during procedures performed on the Processor SP caused by Tips in Tip Holder Assemblies for the following products and lots of Extraction Trays.

    • Terminated Z-2736-2014 FDA Class II initiated 17 Jun 2013

    There is a specific Extraction Tray lot containing Tips that may slightly increase the occurrence of Process SP Processing Error when used due to the Processor SP experiencing difficulty releasing the Tip to the final location in the Tip Assembly. The difficult releasing the Tip can result in a Processing Error and occasionally the Tip to be dropped into the Processor SP.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
PEN
Device Class
Class II
Regulation
21 CFR Part 866.3365
Advisory Committee
Microbiology
FDA Decision
SESE Substantially Equivalent