VerSys® 7841-14-30
by Zimmer, Inc.
VerSys 7841-14-30 is a device of GMDN type press-fit femoral stem prosthesis (FDA Class II; FDA product code LPH, Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented) made by Zimmer, Inc., cleared under FDA 510(k) K953337 on 22 Jan 1996 and listed in FDA GUDID since 24 Nov 2017 (UDI-DI 00889024136144), active as of 23 Oct 2026. 14 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00889024136144 | Date24 Nov 2017 | StatusListed |
| United States | FDA 510(k)SESE | NumberK953337 | Date22 Jan 1996 | StatusActive |
Official records
- AccessGUDID record 00889024136144
- FDA 510(k) database K953337
From the FDA record
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied sterile
Across registries
- FDA 510(k) K953337 decided 22 Jan 1996 Active
- FDA GUDID UDI-DI 00889024136144 listed 24 Nov 2017 Listed
Recalls
-
- Terminated Z-0722-2018
The products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier.
-
- Terminated Z-2580-2017
- Terminated Z-2578-2017
- Terminated Z-2581-2017
- Terminated Z-2579-2017
- Terminated Z-2577-2017
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
-
- Terminated Z-1127-2016
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
-
- Terminated Z-0849-2016
LDPE bag containing the implant adheres to the highly polished implant surface.
-
- Terminated Z-1034-2014
- Terminated Z-1031-2014
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
-
- Terminated Z-0070-2014
The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.
-
- Terminated Z-2290-2010
- Terminated Z-2291-2010
Zimmer, Inc. has determined that there is a potential for a package to be labeled as a 28 mm femoral head but contain a 32 mm femoral head. In addition, there is a potential for the patient record label to be incorrect.
-
- Terminated Z-0198-2010
Manufacturing material may be present on the surface of the implant.
-
- Terminated Z-1066-05
May be mislabeled as to dimensions; may be 12/14 taper, 28mm dia., neck length +3.5mm.
Timeline
- Recall initiated FDA recalls Z-1031-2014
- Recall initiated FDA recalls Z-1127-2016
- Recall initiated FDA recalls Z-0849-2016
- Recall initiated FDA recalls Z-2580-2017
- Recall initiated FDA recalls Z-2578-2017
- Recall initiated FDA recalls Z-2581-2017
- Recall initiated FDA recalls Z-2579-2017
- Recall initiated FDA recalls Z-2577-2017
- Recall initiated FDA recalls Z-0722-2018
- Listed FDA GUDID 00889024136144
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- VerSys®
- Generic name
- Press-fit femoral stem prosthesis
- Model / reference
- 7841-14-30
- Primary DI / UDI-DI
- 00889024136144
- 510(k) Number
- FDA product code
- LPH
- GMDN term
- Press-fit femoral stem prosthesis
- Device class
- Class II
- Regulation number
- 888.3358
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code LPH (PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED).
- Mpact Acetabular System 01.32.6555 by Medacta International SA
- USTAR II Knee System 2915-2060 by UNITED ORTHOPEDIC CORP.
- S-ROM FEMORAL HIP STEM - SIZE 14 X 8 by Joint Medical Products Corp.
- Mpact 3D Metal Foam 25070.2856HC by Medacta International SA
- "UNITED" USTAR II Knee System 2315-1242 by UNITED ORTHOPEDIC CORP.
- CONSENSUS TAPERSET HIP SYSTEM by CONSENSUS ORTHOPEDICS, INC.
- Pantheon System Bridging Collar 0201296 by ADLER ORTHO SPA
- G7® 010002668 by Biomet Orthopedics, Inc.
- METAL ACETABULAR COMPONENT by Encore Orthopedics, Inc.
- Revision Femoral Head 5010.42.053 by Limacorporate