Walkaway TM 40
510(k) K911400
Walkaway TM 40 is a medical device (FDA Class II; FDA product code LRG, Instrument For Auto Reader & Interpretation Of Overnight Suscept. Systems) made by Baxter Healthcare Corp, cleared under FDA 510(k) K911400 on 19 Nov 1991, active as of 11 Oct 2026. 11 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
6 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK911400 | Date19 Nov 1991 | StatusActive |
Official records
- FDA 510(k) database K911400
From the FDA record
- Received by the FDA
- 29 Mar 1991
- FDA decision
- 19 Nov 1991 (Substantially Equivalent)
- FDA review time
- 235 days from receipt to decision
Recalls
-
- Ongoing Z-1088-2022
- Ongoing Z-1083-2022
- Ongoing Z-1086-2022
- Ongoing Z-1084-2022
- Ongoing Z-1085-2022
- Ongoing Z-1087-2022
WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
-
- Terminated Z-1373-2021
- Terminated Z-1372-2021
- Terminated Z-1370-2021
- Terminated Z-1371-2021
Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.
-
- Terminated Z-1990-2014
Springs contained in the access door hinge assembly on the Walk Away Plus instruments are becoming detached causing premature failure of the hinge assembly and a potential safety issue for the customers.
Timeline
- Recall initiated FDA recalls Z-1373-2021
- Recall initiated FDA recalls Z-1372-2021
- Recall initiated FDA recalls Z-1370-2021
- Recall initiated FDA recalls Z-1371-2021
- Recall initiated FDA recalls Z-1088-2022
- Recall initiated FDA recalls Z-1083-2022
- Recall initiated FDA recalls Z-1086-2022
- Recall initiated FDA recalls Z-1084-2022
- Recall initiated FDA recalls Z-1085-2022
- Recall initiated FDA recalls Z-1087-2022
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- LRG
- Device class
- Class II
- Regulation
- 21 CFR Part 866.1640
- Advisory committee
- Microbiology
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code LRG (Instrument For Auto Reader & Interpretation Of Overnight Suscept. Systems).
- MICROSCAN MICROSTREP PLUS PANEL TETRACYCLINE (0.006 TO 16 MCG/ML) by Dade Behring, Inc.
- MICROSCAN MICROSTREP PLUS PANEL MEROPENEM (0.03-4MCG/ML) by Dade Behring, Inc.
- SENSITITRE 18-24 HOURS SUSCEPTIBILITY PLATES by Trek Diagnostic Systems, Inc.
- Sensititre GN7F by Trek Diagnostic Systems, Inc.
- MICROSCAN DRI GRAM POSIT/GRAM NEGA MIC/COMBO PANEL by Baxter Diagnostics, Inc.
- MICROSCAN(R) MICROSTREP PLUS(R) by Siemens Healthcare Diagnostics
- READAR by Alamar Biosciences, Inc.
- Aladin (TM) by Analytical Products, Inc.
- MicroScan Dried Gram Negative MIC/Combo Panels, ESBL Screen by Dade Microscan, Inc.
- AUTOSCEPTOR by Bd Becton Dickinson Vacutainer Systems Preanalytic