Skip to content

Xio Rtp System

by Computerized Medical Systems, Inc. 2 variants

510(k) K102216

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

7 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK102216 Date1 Oct 2010ClearedStatusActive

Recalls

7 recalls · FDA

    • Terminated Z-0130-2016 FDA Class II initiated 31 Aug 2015

    Incorrect Treatment Delivery Using Third Party Fixed Wedges.

    • Terminated Z-1197-2014 FDA Class II initiated 17 Feb 2014

    Dose and Monitor Unit values are not computed correctly when Elekta Motorized Wedges are used with Dynamic Conformal arcs.

    • Terminated Z-0780-2014 FDA Class II initiated 25 Nov 2013

    When the number of fractions is changed for a proton spot beam, the MU are recomputed and rounded if necessary, but the dose is not recalculated as it should be even after saving and re-opening the plan.

    • Terminated Z-0928-2014 FDA Class II initiated 25 Nov 2013

    When bolus is present, the Effective Depth (with bolus) to the weight point returned on the Source Data Report is incorrect.

    • Terminated Z-1644-2012 FDA Not yet classified initiated 30 Apr 2012

    XiO will underestimate the dose delivered when the closed leaf gap is set to 0mm on XiO and a value other than 0mm is set on the accelerator.

    • Terminated Z-2919-2011 FDA Unclassified (pre-2012) initiated 9 Jun 2011

    The XiO Software currently prompts for entry of CT to ED data for a studyset. No indication is given that the Electron Density data should only be used for Photon and Electron calculations and either Relative Stopping Power or Mass Density should be used for Proton calculations (depending on the algorithm).

    • Terminated Z-2582-2011 FDA Unclassified (pre-2012) initiated 16 Nov 2010

    XiO: When a new electron beam is added to a plan, the calculation Algorithm is set to Pencil Beam, and an Aperture is added to the beam, the user can select Port-Port Properties and edit the Material Thickness values. If the algorithm is then changed to Monte Carlo, the calculation will proceed and the user is not warned the Material/Thickness values just entered are not used in the dose calculati

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
MUJ
Device Class
Class II
Regulation
21 CFR Part 892.5050
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent