510(k) K133259

Ysio Max is a medical device (FDA Class II; FDA product code KPR, System, X-Ray, Stationary) made by Siemens Medical Solutions USA, Inc., cleared under FDA 510(k) K133259 on 24 Jan 2014, active as of 22 Oct 2026. 7 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

7 closed

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK133259 Date24 Jan 2014ClearedStatusActive

Official records

From the FDA record

Received by the FDA
23 Oct 2013
FDA decision
24 Jan 2014 (Substantially Equivalent)
FDA review time
93 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K133259 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

7 recalls · FDA

    • Terminated Z-1339-2020 FDA Class II initiated 8 Jan 2020

    A software bug may lead to one image to be assigned to two different patients, which could potentially affect medical diagnosis.

    • Terminated Z-2758-2020 FDA Class II initiated 8 Jan 2020

    Two software issues (1) Using the override function in case of blocked system movements affecting Luminos dRF Max, Luminos Agile Max and Uroskop Omnia Max systems may cause collision of the system under operator control with obstacles or persons .(2) Incorrectly assigned image affecting Ysio Max, Luminos dRF Max and Luminos Agile Max systems with detector MAX Static causing incorrect base for diagnosis

    • Terminated Z-0300-2019 FDA Class II initiated 6 Aug 2018

    In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety lock. This could lead to overloading the safety rope, which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning.

    • Terminated Z-0015-2017 FDA Class II initiated 1 Aug 2016

    Siemens became aware of a potential issue with the patient table for Ysio Max and Multix Fusion systems. There is an unlikely risk that an internal part of the table may detach causing the table top to tilt to the left side which could result in an injury to a patient on the table.

    • Terminated Z-1667-2015 FDA Class II initiated 13 Apr 2015

    Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.

    • Terminated Z-0994-2015 FDA Class II initiated 15 Dec 2014

    Potential for composed images to be flipped before being sent to PACS on systems with software version VE10 and SmartOrtho license. Flipped images may be reversed so the associated annotations, e.g. labels (R/L), may be displayed incorrectly, potentially leading to misdiagnosis.

    • Terminated Z-0002-2015 FDA Class II initiated 2 Sep 2014

    Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images

Timeline

8 dated events

  1. Cleared FDA 510(k) K133259
  2. Recall initiated FDA recalls Z-0002-2015
  3. Recall initiated FDA recalls Z-0994-2015
  4. Recall initiated FDA recalls Z-1667-2015
  5. Recall initiated FDA recalls Z-0015-2017
  6. Recall initiated FDA recalls Z-0300-2019
  7. Recall initiated FDA recalls Z-1339-2020
  8. Recall initiated FDA recalls Z-2758-2020

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
KPR
Device class
Class II
Regulation
21 CFR Part 892.1680
Advisory committee
Radiology
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code KPR (System, X-Ray, Stationary).

Other devices from Siemens Medical Solutions USA, Inc.