Skip to content

Ziehm Solo

by Ziehm Imaging, Inc.

510(k) K092438

Ziehm Solo is a Class II device made by Ziehm Imaging, Inc., cleared under FDA 510(k) K092438 on 23 Dec 2009, active as of 22 Oct 2026. 3 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK092438 Date23 Dec 2009ClearedStatusActive

Official records

Recalls

4 recalls · 1 ongoing · FDA

    • Ongoing Z-0573-2025 FDA Class II initiated 8 Aug 2024

    Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit.

    • Terminated Z-2643-2016 FDA Class II initiated 15 Apr 2016

    Malfunction of the radiation-warning timer and the 3D-specific interval warning. No 5-minute timer dialogs or beeper will occur until the unit is rebooted.

    • Terminated Z-0563-2014 FDA Class II initiated 21 Nov 2013

    Engineering change in 2006 resulted in a minor non-conformity of their devices. Specifically the emergency switch was changed from a standard component to just an option on the devices.

    • Terminated Z-1430-2012 FDA Unclassified (pre-2012) initiated 29 Apr 2011 on shared 510(k) K092438

    On 04/29/2011, Ziehm Imaging, Inc., Orlando, FL recalled the double foot switch, type KF 2 1S/1S-MED-AP, a component used with Ziehm Imaging GmbH Mobile C-arm due to the potential of unintentional continued activation of fluoroscopic X-ray as the foot switch may not become inactive when the pedal is released.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

5 dated events

  1. Cleared FDA 510(k) K092438
  2. Recall initiated FDA recalls Z-1430-2012
  3. Recall initiated FDA recalls Z-0563-2014
  4. Recall initiated FDA recalls Z-2643-2016
  5. Recall initiated FDA recalls Z-0573-2025

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
OWB
Device class
Class II
Regulation
21 CFR Part 892.1650
Advisory committee
Radiology
FDA decision
SESE Substantially Equivalent

Devices cleared under K092438

1 device · FDA GUDID

Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.

Devices cleared under K092438
UDI-DISize and differences
Ziehm Solo
Ziehm Imaging GmbH
SoloEZIEZIEHMSOLO1 —No longer on the market

Similar devices

Other devices filed under FDA product code OWB (Interventional Fluoroscopic X-Ray System).

Other devices from Ziehm Imaging, Inc.

See all devices from Ziehm Imaging, Inc. →