Ziehm Vision 2
510(k) K073346
Ziehm Vision 2 is a Class II device made by Ziehm Imaging, Inc., cleared under FDA 510(k) K073346 on 21 Mar 2008, active as of 22 Oct 2026. One FDA recall is linked to it.
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK073346 | Date21 Mar 2008 | StatusActive |
Official records
- FDA 510(k) database K073346
Recalls
-
- Ongoing Z-0575-2025
Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit.
-
- Terminated Z-2644-2016
Malfunction of the radiation-warning timer and the 3D-specific interval warning. No 5-minute timer dialogs or beeper will occur until the unit is rebooted.
-
- Terminated Z-0565-2014
- Terminated Z-0564-2014
Engineering change in 2006 resulted in a minor non-conformity of their devices. Specifically the emergency switch was changed from a standard component to just an option on the devices.
-
- Terminated Z-1430-2012
On 04/29/2011, Ziehm Imaging, Inc., Orlando, FL recalled the double foot switch, type KF 2 1S/1S-MED-AP, a component used with Ziehm Imaging GmbH Mobile C-arm due to the potential of unintentional continued activation of fluoroscopic X-ray as the foot switch may not become inactive when the pedal is released.
Timeline
- Cleared FDA 510(k) K073346
- Recall initiated FDA recalls Z-1430-2012
- Recall initiated FDA recalls Z-0565-2014
- Recall initiated FDA recalls Z-0564-2014
- Recall initiated FDA recalls Z-2644-2016
- Recall initiated FDA recalls Z-0575-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- JAK
- Device class
- Class II
- Regulation
- 21 CFR Part 892.1750
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Devices cleared under K073346
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| Ziehm Vision2 30355 | 30355 | EZIEZIEHMVISION1 | — | No longer on the market |
| Ziehm Vision2 FD 30208 | 30208 | EZIEZIEHMVISIONFD1 | — | On the market |
| Ziehm Vision2 FD Vario 30239 | 30239 | EZIEZIEHMVISIONFDV1 | — | No longer on the market |
| Ziehm Vision2 FD Vario 3D 30238 | 30238 | EZIEZIEHMVISIONFDV3D1 | — | No longer on the market |
| Ziehm Vision2 Vario 30237 | 30237 | EZIEZIEHMVISIONV1 | — | No longer on the market |
| Ziehm Vision2 Vario 3D 30236 | 30236 | EZIEZIEHMVISIONV3D1 | — | No longer on the market |
Similar devices
Other devices filed under FDA product code JAK (System, X-Ray, Tomography, Computed).
- ITERATIVE RECONSTRUCTION TECHNIQUE SOFTWARE APPLICATION by Philips Healthcare (Cleveland)
- CardIQ Xpress 2.0 with Snapshot Freeze Option CardIQ Xpress 2.0 Reveal by GE Medical Systems SCS
- Al-Rad Companion (Pulmonary) by SIEMENS MEDICAL SOLUTIONS USA, INC.
- MYOPERFUSION by Toshiba Medical Systems
- SOMATOM SMILE by SIEMENS MEDICAL SOLUTIONS USA, INC.
- SCAN BAND by Cook, Inc.
- HEAD-SPINE STABILIZER HSS-1 by Imaging Accessories, Inc.
- ADVANCE VESSEL ANALYSIS II by GE Healthcare
- VIATRONIC V3D CALCIUM SCORING by Viatronix, Inc.
- PRIMELUNG by Accuimage Diagnostics Corp.