Ziehm Vision2 FD 30208
Ziehm Vision2 FD 30208 is a Class II device made by Ziehm Imaging GmbH, cleared under FDA 510(k) K073346 on 21 Mar 2008 and listed in FDA GUDID since 23 Sep 2016 (UDI-DI EZIEZIEHMVISIONFD1), active as of 12 Nov 2026. One FDA recall is linked to it.
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | NumberEZIEZIEHMVISIONFD1 | Date23 Sep 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK073346 | Date21 Mar 2008 | StatusActive |
Official records
- FDA 510(k) database K073346
Across registries
- FDA 510(k) K073346 decided 21 Mar 2008 Active
- FDA GUDID UDI-DI EZIEZIEHMVISIONFD1 listed 23 Sep 2016 Listed
Recalls
-
- Ongoing Z-0575-2025
Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit.
-
- Terminated Z-2644-2016
Malfunction of the radiation-warning timer and the 3D-specific interval warning. No 5-minute timer dialogs or beeper will occur until the unit is rebooted.
-
- Terminated Z-0565-2014
- Terminated Z-0564-2014
Engineering change in 2006 resulted in a minor non-conformity of their devices. Specifically the emergency switch was changed from a standard component to just an option on the devices.
-
- Terminated Z-1430-2012
On 04/29/2011, Ziehm Imaging, Inc., Orlando, FL recalled the double foot switch, type KF 2 1S/1S-MED-AP, a component used with Ziehm Imaging GmbH Mobile C-arm due to the potential of unintentional continued activation of fluoroscopic X-ray as the foot switch may not become inactive when the pedal is released.
Timeline
- Cleared FDA 510(k) K073346
- Recall initiated FDA recalls Z-1430-2012
- Recall initiated FDA recalls Z-0565-2014
- Recall initiated FDA recalls Z-0564-2014
- Recall initiated FDA recalls Z-2644-2016
- Listed FDA GUDID EZIEZIEHMVISIONFD1
- Recall initiated FDA recalls Z-0575-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Ziehm Vision2 FD
- Generic name
- Mobile general-purpose fluoroscopic x-ray system, digital
- Model / reference
- 30208
- Primary DI / UDI-DI
- EZIEZIEHMVISIONFD1
- 510(k) Number
- FDA product code
- JAK
- GMDN term
- Mobile general-purpose fluoroscopic x-ray system, digital
- Device class
- Class II
- Regulation number
- 892.1750
- GUDID distribution status
- In Commercial Distribution