Is this your brand? Sign in to claim ownership.

Sign in to claim this brand

AESCULAP, INC.

US

Last updated September 17, 2026

Part of Aesculap AG

Information

Country
US
Address
3773 CORPORATE PKWY., CENTER VALLEY, PA, 18034
Website
bbraun.com
LinkedIn
—
Facebook
—
Phone
+1(800)282-9000
PRRC Contact
—
EUDAMED SRN
—
FDA FEI Number
2916714
DUNS Number
082594540

Catalogue (15470)

Page 15 of 310
DeviceModel / ReferenceRegistriesClassStatus
Aesculap MG887R
FDA UDI
Class IActive
Aesculap MG640R
FDA UDI
Class IActive
Aesculap OG128R
FDA UDI
Class IUnknown
Prospace® SN024P
FDA UDI
Class IIUnknown
S4 Cervical SW144T
FDA UDI
Class IIActive
Aesculap MN228R
FDA UDI
Class IIActive
Aesculap FG478T
FDA UDI
Class IIActive
Aesculap PM618R
FDA UDI
Class IIActive
Aesculap MA002R
FDA UDI
Class IActive
Aesculap MA501R
FDA UDI
Class IIActive
Progav 2.0 FX606T
FDA UDI
Class IIActive
Progav 2.0 FX382T
FDA UDI
Class IIActive
Ennovate SZ297R
FDA UDI
Class IActive
Aesculap MB380R
FDA UDI
Class IIActive
Columbus NE379R
FDA UDI
Class IUnknown
Aesculap FF665R
FDA UDI
Class IActive
Aesculap LX210R
FDA UDI
Class IActive
Columbus® UC NN440
FDA UDI
Class IIActive
Susi BH111SU
FDA UDI
Class IUnknown
Progav 2.0 FX653T
FDA UDI
Class IIActive
Elan 4 MD756
FDA UDI
Class IActive
Aesculap MA508R
FDA UDI
Class IUnknown
Aesculap MA588R
FDA UDI
Class IActive
A-Space Sibd SJ632P
FDA UDI
Class IIUnknown
Macs Tl SX827T
FDA UDI
Class IIActive
Sse SX100T
FDA UDI
Class IIUnknown
Aesculap PM612R
FDA UDI
Class IIActive
Aesculap BA723
FDA UDI
Class IActive
Aesculap FK724R
FDA UDI
Class IActive
Atractor FC511R
FDA UDI
Class IActive
Hydrolift SV005T
FDA UDI
Class IIUnknown
Aesculap FK672R
FDA UDI
Class IActive
Prospace® SN013P
FDA UDI
Class IIUnknown
Aesculap DE631R
FDA UDI
Class IUnknown
Tspace® Peek FJ693P
FDA UDI
Class IIActive
Excia® NK690K
FDA UDI
Class IIUnknown
S4 Cervical SX465T
FDA UDI
Class IIActive
Aesculap EK110SU
FDA UDI
Class IIActive
Aesculap BH183R
FDA UDI
Class IActive
VEGA System® PS NX101
FDA UDI
Class IIActive
Atraumata FC323T
FDA UDI
Class IActive
Aesculap EK051R
FDA UDI
Class IIActive
Enduro NB027K
FDA UDI
Class IIUnknown
Aesculap MG643R
FDA UDI
Class IActive
Tspace ME927R
FDA UDI
Class IActive
Diadust FM585R
FDA UDI
Class IActive
Aesculap FL401R
FDA UDI
Class IActive
Atraumata FB734R
FDA UDI
Class IActive
Aesculap NR442K
FDA UDI
Class IIUnknown
Aesculap PO141R
FDA UDI
Class IIActive

Related Companies

Authorised Representatives

No authorised representatives on record.

Importers

Not available yet.

Notified Bodies

Not available yet.

Authorities

Not available yet.

FDA Recalls (44)

Recalls reported by this firm to the FDA. Listed at company level — FDA recall records don't identify a specific catalogue device.

DateRecall No.ClassStatusReason
Jan 15, 2026Z-1485-2026—open, classified

There is the potential for the length of the trocar shaft to be too long.

Sep 24, 2025Z-0311-2026—open, classified

It was determined there is the potential of silicone fragments detaching from the yellow sealing component.

Sep 24, 2025Z-0310-2026—open, classified

It was determined there is the potential of silicone fragments detaching from the yellow sealing component.

Sep 24, 2025Z-0309-2026—open, classified

It was determined there is the potential of silicone fragments detaching from the yellow sealing component.

Sep 24, 2025Z-0312-2026—open, classified

It was determined there is the potential of silicone fragments detaching from the yellow sealing component.

Jan 31, 2025Z-1308-2025—open, classified

The forceps have been used in ways not covered by the design resulting in breakage of the clamps.

Jan 31, 2025Z-1309-2025—open, classified

The forceps have been used in ways not covered by the design resulting in breakage of the clamps.

Jan 31, 2025Z-1310-2025—open, classified

The forceps have been used in ways not covered by the design resulting in breakage of the clamps.

Jan 31, 2025Z-1312-2025—open, classified

The forceps have been used in ways not covered by the design resulting in breakage of the clamps.

Jan 31, 2025Z-1311-2025—open, classified

The forceps have been used in ways not covered by the design resulting in breakage of the clamps.

May 13, 2024Z-2139-2024—open, classified

Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.

Jan 9, 2024Z-1246-2024—open, classified

The sterile blister packaging may be damaged, and sterility may be compromised.

Jan 9, 2024Z-1244-2024—open, classified

The sterile blister packaging may be damaged, and sterility may be compromised.

Jan 9, 2024Z-1245-2024—open, classified

The sterile blister packaging may be damaged, and sterility may be compromised.

Jan 9, 2024Z-1247-2024—open, classified

The sterile blister packaging may be damaged, and sterility may be compromised.

Jan 9, 2024Z-1243-2024—open, classified

The sterile blister packaging may be damaged, and sterility may be compromised.

Dec 29, 2023Z-0866-2024—open, classified

Needle Holder MB215R was inadvertently distributed as MB362R. Similarly, MB362R was distributed as MB215R during the affected distribution date range. The issue may lead to a short delay in procedure.

Dec 29, 2023Z-0867-2024—open, classified

Needle Holder MB215R was inadvertently distributed as MB362R. Similarly, MB362R was distributed as MB215R during the affected distribution date range. The issue may lead to a short delay in procedure.

Nov 30, 2015Z-1043-2016—terminated

Aesculap Inc. US has initiated a recall on Tibial and Femur extension sterile pressfit stem implant packaging that are used for Columbus Revision/Enduro Knee Implants because it is labeled as Cementless. US product is indicated in the US for use with Bone Cement only.

Nov 17, 2015Z-0566-2016—terminated

AAG received a customer complaint that corrosion had been found by a hospitals central sterile service department on the ball joint of an FC424R blade. Investigation revealed that the corrosion was due to incorrect material used in the manufacturing of the disk spring of the ball joint by the supplier

Nov 10, 2015Z-0448-2016—terminated

The flexible shaft of the instruments may break during orthopedic spinal surgery. There is a risk of a delay in surgery or that a fragment from the shaft could be left in a patient.

Nov 10, 2015Z-0424-2016—terminated

The flexible shaft of the instruments may break during orthopedic spinal surgery. There is a risk of a delay in surgery or that a fragment from the shaft could be left in a patient.

Nov 10, 2015Z-0423-2016—terminated

The flexible shaft of the instruments may break during orthopedic spinal surgery. There is a risk of a delay in surgery or that a fragment from the shaft could be left in a patient.

Nov 10, 2015Z-0422-2016—terminated

The flexible shaft of the instruments may break during orthopedic spinal surgery. There is a risk of a delay in surgery or that a fragment from the shaft could be left in a patient.

Jul 7, 2015Z-2326-2015—terminated

A FW103R torque wrench that was improperly reassembled after reprocessing caused a screw to break during implantation due to too much torque. The firm did not include instructions for reassembly after reprocessing.

Dec 17, 2014Z-1012-2015—terminated

Some Miethke Shunt System accessories may have an incorrect MR safety symbol on the shelf box label. The label should have NO symbol but instead is marked with the MR conditional symbol. The product is MR unsafe. This information is only required to be in the Instructions for Use (IFU).

Nov 11, 2014Z-0814-2015—terminated

Aesculap Inc. (AIC (USA)) initiated a recall of Miethke Shunt System accessories due to missing patients' labels for the miniNAV (FV660T). There are no patient injuries resulting from this issue.

Oct 24, 2014Z-0502-2015—terminated

Aesculap Inc. (AIC (USA)) initiated a recall of Miethke Shunt System, Dual Switch Valve, due to incorrect product labeling on three units. The incorrect product labeling was the wrong expiration date, 2024-09, on the inner labeling of the sterile bag. However, the expiration date on the outer box was correct. No patient injuries were reported as a result of this issue.

Jun 13, 2014Z-2024-2014—terminated

Aesculap Inc. (AIC (USA)) has initiated a voluntary recall of Dafilon Black 10/0 (0.2) 15 cm DRm4 non-sterile non-absorbable polyamide Surgical Suture due to a labeling issue. Box labels do state non-sterile; however there is an additional label that is affixed on the box that states the product is sterile.

Apr 21, 2014Z-1676-2014—terminated

AIC (USA) received information regarding 3 complaints in which the deflector did not move freely on the ventricular catheter.

Dec 23, 2013Z-0989-2014—terminated

The accuracy is out of specification for the Aesculap proSA Adjustment Disc Size L.

Dec 9, 2013Z-0623-2014—terminated

Certain lots of the Aesculap Hi Line XXS handpiece (part # GB790R) were recalled due to a possible failure of the handpiece.

Sep 20, 2013Z-0361-2014—terminated

The trial insertion instrument may disassemble during use due to a potentially weak weld on then trial inserter shaft block area located near the thumb wheel.

Sep 19, 2013Z-0492-2014—terminated

The Columbus Revision CoCr 15 mm distal and posterior implant spacers, sizes F5-F7, were not compatible.

Sep 10, 2013Z-0479-2014—terminated

Complaints were received for the GN161 Bipolar Foot Control reporting that the bipolar energy did not stop after release of the foot control.

Jul 16, 2013Z-0031-2014—terminated

Changes were implemented to the UNITRAC instructions for use to ensure safe use and potentially extend the life of the product. This included recommending the use of Sterile Cover JG901 and new Quick Release Coupling RT02OR to eliminate the need to sterilize the UNITRAC and a functional check was added prior to use to ensure optimal functionality of the UNITRAC system.

Jun 18, 2013Z-1806-2013—terminated

Aesculap Implant Systems, LLC initiated a recall of the Aesculap Prospace PEEK 5 degree Implant - 8 x 8.5 x 22 mm (SN038P) because the device is incorrectly etched with an 8 degree lordosis instead of a 5 degree lordosis. The device is labeled correctly, only the etching of the implant may be incorrect.

Jan 18, 2013Z-0887-2013—terminated

The bipolar energy did not stop after release of the foot pedal.

Jun 4, 2012Z-0805-2013—terminated

The deflector may not move freely on the ventricular catheter. The deflector comes pre-positioned at the 5 cm mark on the catheter. In some cases the surgeon will want to adjust the deflector to a different location on the catheter before placing it in the ventricle. There is a small percentage of cases where aggressively attempting to adjust the deflector has a potential to cause the catheter

May 9, 2012Z-0082-2013—terminated

The manufacturer informed the recalling firm of a potential laser marking error in size identification on two Columbus 4-IN-1 Femoral Cutting Guide Instruments. The guides are correctly dimensioned, but the laser markings have been incorrectly etched.

Jul 19, 2011Z-3014-2011—terminated

Aesculap Implant Systems received a notification from the manufacturer, Aesculap, AG of a fractured CoCr neck adapter.

Nov 22, 2010Z-1474-2011—terminated

Potential safety issue in that the control box can misinterpret the information sent from the low speed motor if the control magnet inside the motor changes its posthion.

Nov 1, 2007Z-0709-2008—terminated

Breakage: the inner rod of the applier may corrode and break

Nov 1, 2007Z-0708-2008—terminated

Breakage: the inner rod of the applier may corrode and break