Is this your brand? Sign in to claim ownership.

Sign in to claim this brand

Trilliant Surgical LLC

US

Last updated September 17, 2026

Information

Country
US
Address
727 N. Shepherd Dr, Houston, TX, 77007
Website
—
LinkedIn
—
Facebook
—
Email
—
Phone
—
PRRC Contact
—
EUDAMED SRN
—
FDA FEI Number
3007420745
DUNS Number
—

Catalogue (1635)

Page 5 of 33
DeviceModel / ReferenceRegistriesClassStatus
Tiger 210-99-101
FDA UDI
Class IActive
Arsenal Ankle 13 Hole Anterolateral Tibia Plate, R 300-11-008
FDA UDI
Class IIActive
Two-Step Hammertoe 204-24-006
FDA UDI
Class IIActive
Tiger 203-40-352
FDA UDI
Class IIActive
2.4mm x 30mm Gridlock Screw 301-24-030
FDA UDI
Class IIActive
Tiger 200-40-050
FDA UDI
Class IIActive
Arsenal Foot 308-22-018
FDA UDI
Class IIActive
Tiger 213-20-002
FDA UDI
Class IActive
Tiger 203-30-146
FDA UDI
Class IIActive
Tiger 200-24-030
FDA UDI
Class IIActive
Tiger 211-24-007
FDA UDI
Class IActive
Smooth Gridlock Screw Washer, Large 303-50-003
FDA UDI
Class IIActive
2.4mm x 28mm Gridlock Screw 301-24-028
FDA UDI
Class IIActive
Tiger 200-30-014
FDA UDI
Class IIActive
Tiger 210-45-007
FDA UDI
Class IActive
Tiger 203-30-012
FDA UDI
Class IIActive
Tiger 203-24-020
FDA UDI
Class IIActive
Tiger 201-20-008
FDA UDI
Class IIActive
Two-Step Hammertoe 214-00-007
FDA UDI
Class IActive
Gridlock Ankle Plate Instrument Module Lid 320-99-113
FDA UDI
Class IActive
Tiger 201-40-354
FDA UDI
Class IIActive
Arsenal Foot 308-35-055
FDA UDI
Class IIActive
Arsenal Foot 330-01-202
FDA UDI
Class IActive
Arsenal Foot 300-82-005
FDA UDI
Class IIActive
8mm/10mm Staple Pilot Drill 510-10-002
FDA UDI
Class IActive
Tiger 215-00-006
FDA UDI
Class IActive
3.0mm x 28mm Gridlock Locking Screw 302-30-028
FDA UDI
Class IIActive
Tiger 201-40-348
FDA UDI
Class IIActive
3.0mm x 18mm Gridlock Locking Screw 302-30-018
FDA UDI
Class IIActive
2 Hole VLC Gridlock Plate 300-50-001
FDA UDI
Class IIActive
Arsenal Foot 300-80-008
FDA UDI
Class IIActive
5.5mm x 4.0mm x 10mm Two-Step Hammer Toe Implant 204-40-210
FDA UDI
Class IIActive
9mm Twist Subtalar Trial 112-10-009
FDA UDI
Class IActive
4.0mm x 36mm Gridlock Ankle Screw 304-40-036
FDA UDI
Class IIActive
6mm Wedge VL Gridlock Plate 300-54-006
FDA UDI
Class IIActive
Arsenal Foot 308-27-024
FDA UDI
Class IIActive
4.0mm x 42mm Gridlock Ankle Syndesmotic Screw LP 306-41-042
FDA UDI
Class IIActive
3.5mm x 40mm Gridlock Ankle Locking Screw 305-35-040
FDA UDI
Class IIActive
Arsenal Foot 330-27-005
FDA UDI
Class IActive
2.4mm x 10mm Gridlock Screw 301-24-010
FDA UDI
Class IIActive
Tiger 200-20-042
FDA UDI
Class IIActive
Tiger 206-55-080
FDA UDI
Class IIActive
2.4mm Gridlock Screw Overdrill bit 310-25-001
FDA UDI
Class IActive
Tiger 207-70-095
FDA UDI
Class IIActive
Gridlock Screw Driver bit 310-30-003
FDA UDI
Class IActive
Tiger 205-70-105
FDA UDI
Class IIActive
Arsenal Foot 308-27-048
FDA UDI
Class IIActive
Tiger 203-24-142
FDA UDI
Class IIActive
Tiger 203-40-254
FDA UDI
Class IIActive
Tiger 212-00-001
FDA UDI
Class IActive

Related Companies

Authorised Representatives

No authorised representatives on record.

Importers

Not available yet.

Notified Bodies

Not available yet.

Authorities

Not available yet.

FDA Recalls (24)

Recalls reported by this firm to the FDA. Listed at company level — FDA recall records don't identify a specific catalogue device.

DateRecall No.ClassStatusReason
Nov 4, 2019Z-1635-2020—terminated

The firm has identified that parts from the kit, lot: TSL007202, have the potential to contain the incorrect contents of the 8mm HTR Sterile Hammer Toe Reaming Kit. The kit should contain (2) reamers total, (1) Concave Reamer and (1) Convex Reamer; there is the potential that kits from the identified lot may have (2) of the same reamer type as opposed to one (1) of each type.

Oct 10, 2019Z-1892-2020—terminated

Labeling Error; specifically, drill bits from lots # TSL008495A are marked with the lot # of TSL008495. This could lead to parts being misidentified and impact device traceability.

Mar 14, 2019Z-0799-2020—terminated

A screwdriver used to facilitate implant insertion and removal was found to be out of specification for the minor diameter of the drive feature causing the driver to be unable to retain screws as intended.

Jul 19, 2018Z-1608-2020—terminated

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Jul 19, 2018Z-1612-2020—terminated

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Jul 19, 2018Z-1614-2020—terminated

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Jul 19, 2018Z-1610-2020—terminated

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Jul 19, 2018Z-1609-2020—terminated

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Jul 19, 2018Z-1605-2020—terminated

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Jul 19, 2018Z-1611-2020—terminated

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Jul 19, 2018Z-1606-2020—terminated

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Jul 19, 2018Z-1607-2020—terminated

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Jul 19, 2018Z-1613-2020—terminated

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Mar 27, 2018Z-2339-2018—terminated

Firm is voluntarily recalling cannulated driver handles (210-00-003 and 210-00-004) that may demonstrate a degradation of function with the adapter collar sticking, and/or not retaining the mating adapter male shaft.

Mar 27, 2018Z-2338-2018—terminated

Firm is voluntarily recalling cannulated driver handles (210-00-003 and 210-00-004) that may demonstrate a degradation of function with the adapter collar sticking, and/or not retaining the mating adapter male shaft.

Dec 8, 2016Z-0960-2017—terminated

Four lots of Tiger Screw Sterilization Tray Base Assemblies were released to the field with a printing error where k-wire description markings indicate diameters of 0.45 and 0.35 instead of 0.045" and 0.035", respectively.

Sep 2, 2016Z-0609-2017—terminated

Screws were identified as having the incorrect part number and lot number marked upon them. The part number incorrectly indicated a length of 115 mm rather than 110 mm. The lot number incorrectly indicated lot number of TSL003765 rather than TSL003721.

Aug 6, 2015Z-2490-2015—terminated

The affected parts subject to the recall are out of specification, resulting in in the loss of the optimal locking screw functionality of the plates. The thread depth of the screw holes is insufficient, potentially limiting the ability to engage and lock with associated mating locking screws.

Jul 31, 2015Z-2738-2015—terminated

Gridlock screw driver bits for the Gridlock Plating System were identified to be out of specification and have the potential for shallow engagement between the driver and the screw, resulting in an increased likelihood of stripping screw heads intraoperatively. Specifically, the minor diameter of the drive feature is above the specification, potentially reducing the engagement between the driver and the screw head.

Jul 20, 2015Z-2452-2015—terminated

The cannulation of the 3.0/4.0 Cannulated Drill Bit inner diamter is too small to allow for the guide wire to pass completely through.

Jul 20, 2015Z-2421-2015—terminated

The cutting efficiency of the Countersinks was out of specification.

Jul 20, 2015Z-2456-2015—terminated

The cannulation of the Concave and Convex Reamer Heads may be too small to allow the Guide Pin to freely pass through.

Jul 20, 2015Z-2464-2015—terminated

Sterilization tray lids for the Tiger and Tiger Headless Cannulated Screw Systems were reported to occasionally be difficult to remove from the sterilization tray.

Apr 14, 2009Z-1438-2010—terminated

Firm received several complaints of field failures, e.g., "broke during surgery" and the firm initiated a design change. A communication to sales staff 04/14/2009 instructed them to recover and reconcile these products from user physicians.