Page 7 of 33
DeviceModel / ReferenceRegistriesClassStatus
Tiger 203-20-022
FDA UDI
Class IIActive
Tiger 203-24-014
FDA UDI
Class IIActive
Tiger 215-00-004
FDA UDI
Class IActive
Tiger 203-40-250
FDA UDI
Class IIActive
3.5mm x 20mm Gridlock Ankle Screw 304-35-020
FDA UDI
Class IIActive
Tiger 203-20-032
FDA UDI
Class IIActive
Tiger 201-24-030
FDA UDI
Class IIActive
Tiger 205-70-001
FDA UDI
Class IIActive
Tiger 211-40-002
FDA UDI
Class IActive
9 Hole 1/3 Tubular Gridlock Ankle Plate 300-63-002
FDA UDI
Class IIActive
Tiger 215-55-003
FDA UDI
Class IActive
4.0mm x 60mm Gridlock Ankle Syndesmotic Screw LP 306-41-060
FDA UDI
Class IIActive
10mm Convex Reamer 220-10-002
FDA UDI
Class IActive
4 Hole VL Gridlock Plate 300-55-003
FDA UDI
Class IIActive
Tiger 203-40-224
FDA UDI
Class IIActive
Tiger 201-30-146
FDA UDI
Class IIActive
Tiger 200-40-048
FDA UDI
Class IIActive
Tiger 201-40-050
FDA UDI
Class IIActive
11mm Disco Subtalar Implant 102-20-011
FDA UDI
Class IIActive
Arsenal Foot 300-82-025
FDA UDI
Class IIActive
Minimally Invasive Bunion Plate, Right 300-70-001
FDA UDI
Class IIActive
3S Hemi 111-00-010
FDA UDI
Class IActive
Tiger 200-45-001
FDA UDI
Class IIActive
Arsenal Foot 308-35-012
FDA UDI
Class IIActive
Arsenal Foot 308-40-048
FDA UDI
Class IIActive
Tiger 200-40-046
FDA UDI
Class IIActive
Arsenal Ankle 340-01-202
FDA UDI
Class IActive
Tiger 201-40-256
FDA UDI
Class IIActive
Tiger 211-01-201
FDA UDI
Class IActive
Tiger 202-20-024
FDA UDI
Class IIActive
Tiger 340-15-007
FDA UDI
Class IActive
3.0mm x 28mm Gridlock Screw 301-30-028
FDA UDI
Class IIActive
Arsenal Foot 307-22-012
FDA UDI
Class IIActive
Arsenal Foot 308-27-014
FDA UDI
Class IIActive
3.0mm x 14mm Gridlock Locking Screw 302-30-014
FDA UDI
Class IIActive
Tiger 210-00-101
FDA UDI
Class IActive
Arsenal Foot 300-81-002
FDA UDI
Class IIActive
Tiger 203-40-214
FDA UDI
Class IIActive
Tiger 203-20-018
FDA UDI
Class IIActive
Tiger 202-30-014
FDA UDI
Class IIActive
2.7mm x 12mm Gridlock Ankle Locking Screw 305-27-012
FDA UDI
Class IIActive
Tiger 213-30-002
FDA UDI
Class IActive
Tiger 200-30-026
FDA UDI
Class IIActive
Tiger 206-70-050
FDA UDI
Class IIActive
Tiger 200-20-022
FDA UDI
Class IIActive
Arsenal Ankle 309-45-110
FDA UDI
Class IIActive
Arsenal Foot 311-02-020
FDA UDI
Class IActive
Subtalar Implant System Sterilization Tray Lid 112-99-011
FDA UDI
Class IActive
4.0mm x 24mm Cannulated Screw 201-40-024
FDA UDI
Class IIActive
Tiger 200-45-003
FDA UDI
Class IIActive

Related Companies

Authorised Representatives

No authorised representatives on record.

Importers

Not available yet.

Notified Bodies

Not available yet.

Authorities

Not available yet.

FDA Recalls (24)

Recalls reported by this firm to the FDA. Listed at company level — FDA recall records don't identify a specific catalogue device.

DateRecall No.ClassStatusReason
Nov 4, 2019Z-1635-2020—terminated

The firm has identified that parts from the kit, lot: TSL007202, have the potential to contain the incorrect contents of the 8mm HTR Sterile Hammer Toe Reaming Kit. The kit should contain (2) reamers total, (1) Concave Reamer and (1) Convex Reamer; there is the potential that kits from the identified lot may have (2) of the same reamer type as opposed to one (1) of each type.

Oct 10, 2019Z-1892-2020—terminated

Labeling Error; specifically, drill bits from lots # TSL008495A are marked with the lot # of TSL008495. This could lead to parts being misidentified and impact device traceability.

Mar 14, 2019Z-0799-2020—terminated

A screwdriver used to facilitate implant insertion and removal was found to be out of specification for the minor diameter of the drive feature causing the driver to be unable to retain screws as intended.

Jul 19, 2018Z-1608-2020—terminated

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Jul 19, 2018Z-1612-2020—terminated

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Jul 19, 2018Z-1614-2020—terminated

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Jul 19, 2018Z-1610-2020—terminated

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Jul 19, 2018Z-1609-2020—terminated

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Jul 19, 2018Z-1605-2020—terminated

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Jul 19, 2018Z-1611-2020—terminated

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Jul 19, 2018Z-1606-2020—terminated

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Jul 19, 2018Z-1607-2020—terminated

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Jul 19, 2018Z-1613-2020—terminated

Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.

Mar 27, 2018Z-2339-2018—terminated

Firm is voluntarily recalling cannulated driver handles (210-00-003 and 210-00-004) that may demonstrate a degradation of function with the adapter collar sticking, and/or not retaining the mating adapter male shaft.

Mar 27, 2018Z-2338-2018—terminated

Firm is voluntarily recalling cannulated driver handles (210-00-003 and 210-00-004) that may demonstrate a degradation of function with the adapter collar sticking, and/or not retaining the mating adapter male shaft.

Dec 8, 2016Z-0960-2017—terminated

Four lots of Tiger Screw Sterilization Tray Base Assemblies were released to the field with a printing error where k-wire description markings indicate diameters of 0.45 and 0.35 instead of 0.045" and 0.035", respectively.

Sep 2, 2016Z-0609-2017—terminated

Screws were identified as having the incorrect part number and lot number marked upon them. The part number incorrectly indicated a length of 115 mm rather than 110 mm. The lot number incorrectly indicated lot number of TSL003765 rather than TSL003721.

Aug 6, 2015Z-2490-2015—terminated

The affected parts subject to the recall are out of specification, resulting in in the loss of the optimal locking screw functionality of the plates. The thread depth of the screw holes is insufficient, potentially limiting the ability to engage and lock with associated mating locking screws.

Jul 31, 2015Z-2738-2015—terminated

Gridlock screw driver bits for the Gridlock Plating System were identified to be out of specification and have the potential for shallow engagement between the driver and the screw, resulting in an increased likelihood of stripping screw heads intraoperatively. Specifically, the minor diameter of the drive feature is above the specification, potentially reducing the engagement between the driver and the screw head.

Jul 20, 2015Z-2452-2015—terminated

The cannulation of the 3.0/4.0 Cannulated Drill Bit inner diamter is too small to allow for the guide wire to pass completely through.

Jul 20, 2015Z-2421-2015—terminated

The cutting efficiency of the Countersinks was out of specification.

Jul 20, 2015Z-2456-2015—terminated

The cannulation of the Concave and Convex Reamer Heads may be too small to allow the Guide Pin to freely pass through.

Jul 20, 2015Z-2464-2015—terminated

Sterilization tray lids for the Tiger and Tiger Headless Cannulated Screw Systems were reported to occasionally be difficult to remove from the sterilization tray.

Apr 14, 2009Z-1438-2010—terminated

Firm received several complaints of field failures, e.g., "broke during surgery" and the firm initiated a design change. A communication to sales staff 04/14/2009 instructed them to recover and reconcile these products from user physicians.