DeviceModel / ReferenceRegistriesClassStatus
Waypoint Stereotactic System K092192
FDA 510(k)
Class IIActive
Microtargeting Guideline 4000 K071364
FDA 510(k)
Class IIActive
VFS1 Bipolar Electrosurgical Generator K200169
FDA 510(k)
Class IIActive
Fhc Microtargeting Drive System K011992
FDA 510(k)
Class IIActive
Micro Targeting Electrode K033173
FDA 510(k)FDA UDI
Class IIActive
Visualase™ Neuro Accessories and Tools 9736545
EUDAMEDFDA UDI
Class IActive
Visualase™ Neuro Accessories and Tools 9736541
EUDAMEDFDA UDI
Class IActive
Visualase™ Neuro Accessories and Tools 9736543
EUDAMEDFDA UDI
Class IActive
Visualase™ Neuro Accessories and Tools 9736537
EUDAMEDFDA UDI
Class IActive
Visualase™ Neuro Accessories and Tools 9736544
EUDAMEDFDA UDI
Class IActive
Visualase™ Neuro Accessories and Tools 9736540
EUDAMEDFDA UDI
Class IActive
Visualase™ Neuro Accessories and Tools 9736539
EUDAMEDFDA UDI
Class IActive
Visualase™ Neuro Accessories and Tools 9736536
EUDAMEDFDA UDI
Class IIaActive
Visualase™ Neuro Accessories and Tools 9736538
EUDAMEDFDA UDI
Class IIaActive

Related Companies

Authorised Representatives

Importers

Not available yet.

Notified Bodies

Not available yet.

Authorities

Not available yet.

FDA Recalls (29)

Recalls reported by this firm to the FDA. Listed at company level — FDA recall records don't identify a specific catalogue device.

DateRecall No.ClassStatusReason
Oct 4, 2024Z-0312-2025—open, classified

A complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of other components was included in the build order for Lot 244517 of P/N 66-IT-06P. Use of the product with a spacer tube which is too long could result in intercranial hemorrhaging and brain damage.

Jun 17, 2020Z-2540-2020—terminated

An electrical sub-circuit was not properly connected. Its purpose is to eliminate the possibility of interference with proper performance of the Guideline 5 Main Processing Unit, if an ESD event (greater than 6kV) were to occur.

May 10, 2018Z-2350-2018—terminated

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

May 10, 2018Z-2349-2018—terminated

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

May 10, 2018Z-2348-2018—terminated

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

May 10, 2018Z-2352-2018—terminated

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

May 10, 2018Z-2354-2018—terminated

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

May 10, 2018Z-2355-2018—terminated

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

May 10, 2018Z-2345-2018—terminated

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

May 10, 2018Z-2358-2018—terminated

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

May 10, 2018Z-2351-2018—terminated

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

May 10, 2018Z-2346-2018—terminated

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

May 10, 2018Z-2359-2018—terminated

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

May 10, 2018Z-2357-2018—terminated

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

May 10, 2018Z-2347-2018—terminated

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

May 10, 2018Z-2353-2018—terminated

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

May 10, 2018Z-2356-2018—terminated

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

Mar 30, 2010Z-0311-2011—terminated

Fixation thumbscrew on the DBS depth stop adapter may be overtightened and damage the implantable lead.

Mar 30, 2010Z-0312-2011—terminated

Fixation thumbscrew on the DBS depth stop adapter may be overtightened and damage the implantable lead.

Mar 30, 2010Z-0313-2011—terminated

Fixation thumbscrew on the DBS depth stop adapter may be overtightened and damage the implantable lead.

Mar 30, 2010Z-0314-2011—terminated

Fixation thumbscrew on the DBS depth stop adapter may be overtightened and damage the implantable lead.

Dec 28, 2009Z-1143-2010—terminated

The raster tab for the GL4K version 1.4 software may assign incorrect depths for waveform recordings and if used to identify the implantation target for the DBS lead, the implant may be placed in the wrong location.

Oct 20, 2008Z-0497-2009—terminated

Mislabeled: May contain an electrode longer than labeled size

Oct 20, 2008Z-0498-2009—terminated

Mislabeled: May contain an electrode longer than labeled size

Oct 20, 2008Z-0499-2009—terminated

Mislabeled: May contain an electrode longer than labeled size

Oct 8, 2008Z-0486-2009—terminated

Measuring fixture is incorrectly graduated.

Oct 31, 2005Z-0180-06—terminated

Power Assist microTargeting System may advance beyond intended target

Oct 31, 2005Z-0179-06—terminated

Power Assist microTargeting System may advance beyond intended target

Mar 14, 2005Z-0895-05—terminated

Potential for non-sterile pin to contaminate sterile field