Micro-X, Ltd.
Active in EUDAMEDAustraliafrom SRN EUDAMED SRN AU-MF-000009037
Micro-X, Ltd. is a medical device manufacturer based in Australia, with 7 devices listed in FDA GUDID, FDA 510(k) and EUDAMED (FDA Class II) and 10 FDA recalls as of 13 Nov 2026.
- Devices
- 7
- Registrations
- EUDAMED
- FDA recalls
- 1010 ongoing
Registry facts
- Registries
- FDA GUDID 5
- FDA 510(k) 2
- EUDAMED 1
- Device classes
- FDA Class II 7
- FDA recalls
- Sites and related companies
- Micro-X Ltd Australia
- Identifiers
- EUDAMED SRN AU-MF-000009037
At a glance
Market reach
- European Union
- EUDAMED actor
- United States
- FDA recalls
Catalogue
Company details
- EUDAMED SRN
- AU-MF-000009037
- EUDAMED status
- Active
- EUDAMED registered name
- Micro-X Ltd
- Registered address
- A14 - 6 MAB, Eastern Promenade, 5042, Adelaide, Australia
- Website
- micro-x.com
- PRRC contact
- Dijana Jerkovic
FDA recalls
Recalls initiated per year
- 2017 0
- 2018 0
- 2019 0
- 2020 0
- 2021 0
- 2022 1
- 2023 2
- 2024 2
- 2025 4
- 2026 1
2026
- 27 Mar 2026 Z-1918-2026 Class II Ongoing X-Ray generator may malfunction resulting in the X-Ray being inoperable.
2025
- 4 Apr 2025 Z-1669-2025 Class II Ongoing Potential for internal fastening mechanism within generator of mobile x-Ray to fail.
- 4 Apr 2025 Z-1670-2025 Class II Ongoing Potential for internal fastening mechanism within generator of mobile x-Ray to fail.
- 28 Mar 2025 Z-1578-2025 Class II Ongoing Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.
- 28 Mar 2025 Z-1579-2025 Class II Ongoing Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.
2024
- 23 Oct 2024 Z-0666-2025 Class II Ongoing During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a very short time such that the orange ring of death is displayed in the tube cooling area of the screen. The system then becomes unresponsive to kV/mAs changes, prep/expose presses. The only way to proceed is to suspend the patient, Exit Nexus DR and relogin and continue the exam.
- 22 Oct 2024 Z-0580-2025 Class II Ongoing Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to 106kPa), Altitude:-102ft to 12,000 ft, (Maximum Gradient:9 degrees Fahrenheit), which mitigate risk of capacitor malfunction, unpredictable behavior, overheating, unintended start-up.
2023
- 28 Nov 2023 Z-0665-2025 Class II Ongoing Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.
- 31 May 2023 Z-0664-2025 Class II Ongoing Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
2022
- 31 Aug 2022 Z-1786-2022 Class II Ongoing Software calibration error with product equip with a Dose Area Product (DAP) meter. This results in the product displaying the incorrect DAP meter values which may impede radiation exposure management decisions
Full distribution text
- 31 Aug 2022 Z-1786-2022 Class II Ongoing Software calibration error with product equip with a Dose Area Product (DAP) meter. This results in the product displaying the incorrect DAP meter values which may impede radiation exposure management decisions
Sites
- Micro-X Ltd AUFEI 3012855798 0 devices
Frequently asked questions
How many devices does Micro-X, Ltd. have in the registry?
The registry currently lists 7 devices for Micro-X, Ltd.. Devices recorded under merged duplicate company entries are counted once.
Which registries list Micro-X, Ltd.'s devices?
Among its 7 most recently registered devices, Micro-X, Ltd. has devices listed in EUDAMED (1 device), FDA 510(k) (2 devices) and FDA GUDID (5 devices). A device can be listed in more than one registry, so these counts overlap.
What risk classes do Micro-X, Ltd.'s devices fall under?
Among its 7 most recently registered devices, the classes recorded for Micro-X, Ltd.'s devices are Class II. Each device page shows its own class and authorisation.
Are there FDA recall records for Micro-X, Ltd.?
FDA recall records list 10 recalls initiated by Micro-X, Ltd., the most recent initiated on 27 Mar 2026. 10 of them are recorded as ongoing. A recall is often a precaution started by the manufacturer; the records say nothing about the company's overall record.
What are the newest devices from Micro-X, Ltd. in the registry?
The most recently registered devices listed for Micro-X, Ltd. include Micro-X Rover and Rover.