Rover
510(k) K211423
Rover is a Class II device made by Micro-X Limited, cleared under FDA 510(k) K211423 on 21 May 2021, active as of 22 Oct 2026. 10 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
10 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK211423 | Date21 May 2021 | StatusActive |
Official records
- FDA 510(k) database K211423
Recalls
-
- Ongoing Z-1918-2026
X-Ray generator may malfunction resulting in the X-Ray being inoperable.
-
- Ongoing Z-1669-2025
- Ongoing Z-1670-2025
Potential for internal fastening mechanism within generator of mobile x-Ray to fail.
-
- Ongoing Z-1579-2025
- Ongoing Z-1578-2025
Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.
-
- Ongoing Z-0666-2025
During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a very short time such that the orange ring of death is displayed in the tube cooling area of the screen. The system then becomes unresponsive to kV/mAs changes, prep/expose presses. The only way to proceed is to suspend the patient, Exit Nexus DR and relogin and continue the exam.
-
- Ongoing Z-0580-2025
Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to 106kPa), Altitude:-102ft to 12,000 ft, (Maximum Gradient:9 degrees Fahrenheit), which mitigate risk of capacitor malfunction, unpredictable behavior, overheating, unintended start-up.
-
- Ongoing Z-0665-2025
Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.
-
- Ongoing Z-0664-2025
Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
-
- Ongoing Z-1786-2022
Software calibration error with product equip with a Dose Area Product (DAP) meter. This results in the product displaying the incorrect DAP meter values which may impede radiation exposure management decisions
Timeline
- Recall initiated FDA recalls Z-1786-2022
- Recall initiated FDA recalls Z-0664-2025
- Recall initiated FDA recalls Z-0665-2025
- Recall initiated FDA recalls Z-0580-2025
- Recall initiated FDA recalls Z-0666-2025
- Recall initiated FDA recalls Z-1579-2025
- Recall initiated FDA recalls Z-1578-2025
- Recall initiated FDA recalls Z-1669-2025
- Recall initiated FDA recalls Z-1670-2025
- Recall initiated FDA recalls Z-1918-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- IZL
- Device class
- Class II
- Regulation
- 21 CFR Part 892.1720
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Devices cleared under K211423
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| Micro-X Rover | MXU-RV19 / Rover | 09357123000013 | — | On the market |
| Micro-X Rover | MXU-RV35 / Rover | 09357123000037 | — | On the market |
| Micro-X Rover | MXU-RV60 / Rover | 09357123000044 | — | On the market |
| Micro-X Rover | MXU-RV71 / Rover | 09357123000051 | — | On the market |
Similar devices
Other devices filed under FDA product code IZL (System, X-Ray, Mobile).
- VMX MOBILE X-RAY SYSTEM by GE MEDICAL SYSTEMS, INC.
- MINXRAY, MODEL HF120/60H POWERPLUS by Minxray, Inc.
- MINXRAY, MODEL HF100H+ by Minxray, Inc.
- ACU-RAY HF SERIES PORTABLE X-RAY UNIT by Sterne Equipment Co., Ltd.
- Portable DR Imaging System by Shantou Institute of Ultrasonic Instruments Co., Ltd. (Siui)
- AMX Navigate Mobile X-Ray System by GE Medical Systems, LLC
- PRACTIX CONVENIO by Philips Medical Systems North America, Inc.
- MOBILEDIAGNOST WDR by Sedecal., Sa.
- iQFlex M Mobile X-ray System, iQFlex MD Mobile X-ray System by FIRST SOURCE INC.
- X-RAY STAND, PORTABLE by General X-Ray, Inc.