Is this your brand? Sign in to claim ownership.

Sign in to claim this brand

Microgenics Corporation

United States·US-MF-000030375

Last updated September 17, 2026

Information

Country
United States
Address
46500 , Kato Road, Fremont, United States
Website
www.thermofisher.com/diagnostics

This website did not respond in time when we last checked. Last checked 2026-09-18.

LinkedIn
—
Facebook
—
Phone
001-510-979-5000
PRRC Contact
Mark Brandes
EUDAMED SRN
US-MF-000030375
FDA FEI Number
3010939897
DUNS Number
—

Catalogue (706)

Page 15 of 15
DeviceModel / ReferenceRegistriesClassStatus
Alinity c DOA MC Qual Calibrator Kit 2 08P6302
EUDAMED
IVD GeneralActive
CEDIA Methamphetamine OFT Assay 10018595
EUDAMEDFDA UDI
IVD GeneralActive
DRI Cocaine Metabolite Assay 0055
EUDAMEDFDA UDI
IVD GeneralActive
Phenobarbital 5P07-21
EUDAMEDFDA UDI
IVD GeneralActive
Acrometrix EV Negative Control 960499
EUDAMEDFDA UDI
IVD GeneralActive
MAS® CardioImmune XL Liquid Assayed Cardiac Marker Control CAI-XLS
EUDAMEDFDA UDI
IVD GeneralActive

Related Companies

Authorised Representatives

Importers

Not available yet.

Notified Bodies

Not available yet.

Authorities

Not available yet.

FDA Recalls (14)

Recalls reported by this firm to the FDA. Listed at company level — FDA recall records don't identify a specific catalogue device.

DateRecall No.ClassStatusReason
Oct 1, 2025Z-0415-2026—open, classified

The firm received complaints from customers reporting vial-to-vial variability when using the control with high sensitivity Troponin I assays. This defect may lead to as delay in patient results for hs Troponin I assays.

Jan 20, 2022Z-1284-2022—open, classified

Due to QMS Tacrolimus Calibrator recovering between 82 to 94% of their target values, resulting in over-quantitation in tacrolimus patient result values when used with the QMS Tacrolimus Immunoassay.

Aug 15, 2016Z-0008-2017—terminated

A reagent stability issue (still under investigation) causes less separation in the calibration curve, which affects results across the entire range. For controls, maximum bias of +/- 35% for low levels, +/- 33% for mid-range levels, and +/- 47% for high levels is observed. Bias in patient samples ranges from -15.7% to +5.2% with an average bias of -5.6% when comparing the QMS everolimus assay to an everolimus LC-MS reference method. Patient samples across the measuring range sub-therapeutic, therapeutic, and supra-therapeutic could be affected.

Jul 25, 2016Z-2884-2016—terminated

Some lots of DRI Salicylate Serum Tox Assay negative patient samples are recovering higher than the product sensitivity claim of 4.4 mg/dL.

Oct 20, 2014Z-0501-2015—terminated

MAS CardioImmune XL, level 1 (lot CXL16011), is showing vial-to-vial variation for the analytes Creatine Kinase-MB (CK-MB) and B-Type Natriuretic Peptide-32 (BNP-32).

Dec 2, 2013Z-1930-2014—terminated

The constituent, Inhibin A, was removed from MAS Omni IMMUNE and MAS Omni IMMUNE PRO controls; however the package insert was not updated to reflect the change.

Dec 2, 2013Z-1929-2014—terminated

The constituent, Inhibin A, was removed from MAS Omni IMMUNE and MAS Omni IMMUNE PRO controls; however the package insert was not updated to reflect the change.

Oct 16, 2013Z-0513-2014—terminated

Microgenics Corp., part of Thermo Fisher Scientific, is recalling CEDIA Tobramycin II Assay (Lot Number: 60169764) due to calibration stability problem relating to two complaints received.

Jul 23, 2013Z-2267-2013—terminated

Surveillance of reagent stability for specific lots of DRI Cocaine Metabolite Reagent show an unexpected decreasing change in the reaction rate, indicating a faster than expected reagent degradation.

May 13, 2013Z-1563-2013—terminated

It has been confirmed from in-house testing that Direct Bilirubin (DBIL), and Total Bilirubin (TBIL) claimed in Thermo Scientific MAS Bilirubin Systems are displaying reduced stability.

Aug 20, 2012Z-0379-2013—terminated

In house testing confirmed that CK-MS and Insulin claimed in both Omni IMMUNE and Omni IMMUNE PRO products are displaying reduced stability.

Dec 14, 2011Z-0637-2012—terminated

Low measurements of Troponin I in the MAS CardioImmune XL Control which were outside the published control ranges.

Nov 11, 2011Z-0773-2012—terminated

Reports that third party control material for 4 lots of Tobramycin Calibrators that are recovering low and outside of the published range.

Jun 8, 2011Z-0297-2012—terminated

Complaint from the sole distributor that the third party control material is recovering high, and outside the published range.