3.0T Laa
Ref 5452049Model 3.0T HNU / UAA / LAA
3.0T Laa (5452049) is a MRI system coil (EU MDR Class I; FDA Class II; FDA product code OUO) made by GE Healthcare Coils, registered in EUDAMED on 12 Mar 2024 (Basic UDI-DI 8406821BUG00195HN), cleared under FDA 510(k) K142098 on 24 Nov 2014 and listed in FDA GUDID since 1 Aug 2016 (UDI-DI 00840682107082), active as of 22 Oct 2026. 3 FDA recalls are linked to it.
Active
Valid until
No date given
FDA recalls
4 ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number8406821BUG00195HN 00840682107082 | Date12 Mar 2024 | StatusActive |
| United States | GUDID listing GUDID | Number00840682107082 | Date1 Aug 2016 | StatusListed |
| United States | FDA 510(k) GUDID | NumberK142098 | Date24 Nov 2014 | StatusActive |
Official records
- EUDAMED public site 8406821BUG00195HN
- AccessGUDID record 00840682107082
- FDA 510(k) database K142098
From the FDA record
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- EUDAMED Basic UDI-DI 8406821BUG00195HN registered 12 Mar 2024 Active
- FDA 510(k) K142098 decided 24 Nov 2014 Active
- FDA GUDID UDI-DI 00840682107082 listed 1 Aug 2016 Listed
Recalls
-
- Ongoing Z-2309-2025
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
-
- Ongoing Z-0520-2025
Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.
-
- Ongoing Z-2466-2024
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
- Ongoing Z-0595-2022
The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.
-
- Terminated Z-3029-2017
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
-
- Terminated Z-0260-2016
When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could exceed IEC60601-2-33 limit of 3.2 W/kg for some scans.
Timeline
- Cleared FDA 510(k) K142098
- Recall initiated FDA recalls Z-0260-2016
- Listed FDA GUDID 00840682107082
- Recall initiated FDA recalls Z-3029-2017
- Recall initiated FDA recalls Z-0595-2022
- Registered EUDAMED 8406821BUG00195HN
- Recall initiated FDA recalls Z-2466-2024
- Recall initiated FDA recalls Z-0520-2025
- Recall initiated FDA recalls Z-2309-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Generic name
- MRI system coil, radio-frequency
- Device name
- 3.0T Hnu / Uaa / Laa
- Model / reference
- 3.0T HNU / UAA / LAA EUDAMED5452049 EUDAMED3.0T LAA GUDID
- Primary DI / UDI-DI
- 00840682107082
- Basic UDI-DI
- 8406821BUG00195HN
- 510(k) Number
- FDA product code
- OUO
- EMDN code
- Z1105018001 Magnetic resonance (MR) radiofrequency surface coils
- GMDN term
- MRI system coil, radio-frequency
- Device class
- Class I EUDAMEDClass II GUDID
- Regulation number
- 892.1200
- EU registration type
- Standard device
- EU market of registration
- France
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Model / reference | 3.0T HNU / UAA / LAA 5452049 | 3.0T LAA |
| Device class | Class I | Class II |
Manufacturer details
- EUDAMED SRN
- US-MF-000001664
- Authorised representative
- GE Medical Systems SCS FR-AR-000000344
Similar devices in this category
Other devices filed under MAGNETIC RESONANCE (MR) RADIOFREQUENCY SURFACE COILS.
- Philips 4598-000-20671 by SHANGHAI CHENGUANG MEDICAL TECHNOLOGIES CO, LTD
- Philips 4598-000-20651 by SHANGHAI CHENGUANG MEDICAL TECHNOLOGIES CO, LTD
- NA 3.0T 16ch AIR AA 5854101 by USA INSTRUMENTS, INC.
- 1.5T Express Coil – 9E Anterior Array 5407990 by GE HEALTHCARE(TIANJIN) COMPANY LIMITED
- 3.0T GE 6-Channel Phased Array Flex Coil 5363393 by GE Medical Systems, LLC
- Vanguard Atlas/Titan 300006366361 by Philips Medical Systems Nederland B.V.
- 1.5T 32Ch Cardiac Coil - Anterior 300007281491 by Philips Medical Systems Nederland B.V.
- Pediatric SPEEDER Stabilizer NC036200 by NEOCOIL, LLC
- NA 1.5T 4CH LARGE FLEX COIL by GE HEALTHCARE(TIANJIN) COMPANY LIMITED
- 3.0T 16ch Shoulder Coil NC143202 by NEOCOIL, LLC