3.0T Uaa
Ref 5486620Model 3.0T HNU / UAA / LAA
3.0T Uaa (5486620) is a MRI system coil (EU MDR Class I; FDA Class II; FDA product code OUO) made by GE Healthcare Coils, registered in EUDAMED on 12 Mar 2024 (Basic UDI-DI 8406821BUG00195HN), cleared under FDA 510(k) K142098 on 24 Nov 2014 and listed in FDA GUDID since 1 Aug 2016 (UDI-DI 00840682107204), active as of 22 Oct 2026. 3 FDA recalls are linked to it.
Active
Valid until
No date given
FDA recalls
4 ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number8406821BUG00195HN 00840682107204 | Date12 Mar 2024 | StatusActive |
| United States | GUDID listing GUDID | Number00840682107204 | Date1 Aug 2016 | StatusListed |
| United States | FDA 510(k) GUDID | NumberK142098 | Date24 Nov 2014 | StatusActive |
Official records
- EUDAMED public site 8406821BUG00195HN
- AccessGUDID record 00840682107204
- FDA 510(k) database K142098
From the FDA record
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- EUDAMED Basic UDI-DI 8406821BUG00195HN registered 12 Mar 2024 Active
- FDA 510(k) K142098 decided 24 Nov 2014 Active
- FDA GUDID UDI-DI 00840682107204 listed 1 Aug 2016 Listed
Recalls
-
- Ongoing Z-2309-2025
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
-
- Ongoing Z-0520-2025
Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.
-
- Ongoing Z-2466-2024
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
- Ongoing Z-0595-2022
The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.
-
- Terminated Z-3029-2017
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
-
- Terminated Z-0260-2016
When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could exceed IEC60601-2-33 limit of 3.2 W/kg for some scans.
Timeline
- Cleared FDA 510(k) K142098
- Recall initiated FDA recalls Z-0260-2016
- Listed FDA GUDID 00840682107204
- Recall initiated FDA recalls Z-3029-2017
- Recall initiated FDA recalls Z-0595-2022
- Registered EUDAMED 8406821BUG00195HN
- Recall initiated FDA recalls Z-2466-2024
- Recall initiated FDA recalls Z-0520-2025
- Recall initiated FDA recalls Z-2309-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Generic name
- MRI system coil, radio-frequency
- Device name
- 3.0T Hnu / Uaa / Laa
- Model / reference
- 3.0T HNU / UAA / LAA EUDAMED5486620 EUDAMED3.0T UAA GUDID
- Primary DI / UDI-DI
- 00840682107204
- Basic UDI-DI
- 8406821BUG00195HN
- 510(k) Number
- FDA product code
- OUO
- EMDN code
- Z1105018001 Magnetic resonance (MR) radiofrequency surface coils
- GMDN term
- MRI system coil, radio-frequency
- Device class
- Class I EUDAMEDClass II GUDID
- Regulation number
- 892.1200
- EU registration type
- Standard device
- EU market of registration
- France
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Model / reference | 3.0T HNU / UAA / LAA 5486620 | 3.0T UAA |
| Device class | Class I | Class II |
Manufacturer details
- EUDAMED SRN
- US-MF-000001664
- Authorised representative
- GE Medical Systems SCS FR-AR-000000344
Similar devices in this category
Other devices filed under MAGNETIC RESONANCE (MR) RADIOFREQUENCY SURFACE COILS.
- Philips dS Base 3.0T 459801290092 by Philips Ds North America LLC
- Philips 4522-133-17141 by SHANGHAI CHENGUANG MEDICAL TECHNOLOGIES CO, LTD
- 1.5T Receive-Only 16ch Flex SPEEDER Medium NC042200 by NEOCOIL, LLC
- Philips 4598-000-51951 by SHANGHAI CHENGUANG MEDICAL TECHNOLOGIES CO, LTD
- Philips dS FootAnkle 16Ch 3.0T 459801295501 by Philips Ds North America LLC
- dS TorsoCardiac 1.5T 459801681991 by Philips Healthcare (Suzhou) Co., Ltd.
- Vanguard Breast MRI Table 453530385162 by Philips Medical Systems Nederland B.V.
- 3.0T 16ch Breast Coil by NEOCOIL, LLC
- GE 8ch Foot Ankle coil 5436920 by GE HEALTHCARE(TIANJIN) COMPANY LIMITED
- Smart Fit TorsoCardiac 1.5T 989710012162 by Philips Healthcare (Suzhou) Co., Ltd.