AccuSITE Needle Guide 610-1018, 610-1031, 610-1124
by Civco Medical Instruments Co., Inc. EUDAMEDGUDID
Active
Ref 610-1018
Regulatory status
Recall ongoing. 1 ongoing FDA recall. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number0841436101TD101B1NA | Date16 May 2026 | StatusActive |
| United States | GUDID listing GUDID | Number10841436101650 | Date22 Aug 2017 | StatusIn distribution |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
- European Union
- United States
Certificates
EUDAMED lists no certificate for this device's Basic UDI-DI.
Recalls
-
- Ongoing Z-1892-2025
During the attachment of the needle guide to the covered probe, the tight tolerance and fit of the needle guide can cause a pinhole breach in the transducer cover when latching the guide in place.
-
- Terminated Z-2888-2018
A cover failed leak testing during validation testing of new needle guides from a replacement tool.
-
- Terminated Z-0286-2016
- Terminated Z-0288-2016
- Terminated Z-0289-2016
- Terminated Z-0290-2016
- Terminated Z-0287-2016
Sterility of the product cannot be assured.
Specification
- Trade Name
- AccuSITE Needle Guide
- Generic Name
- Needle guide, single-use
- Device Name
- General Purpose Guidance – Out-of-Plane, Fixed Angle (Accusite guide)
- Model / Ref / Catalog
- 610-1018, 610-1031, 610-1124
- Primary DI / UDI-DI
- 10841436101650
- Basic UDI-DI
- 0841436101TD101B1NA
- 510(k) Number
- FDA Product Code
- ITX
- EMDN Code
- A018002 Needle guides
- GMDN Term
- Needle guide, single-use
- Device Class
- Class IIa EUDAMEDClass II GUDID
- Regulation Number
- 892.1570
- Market of Registration
- Germany
- Commercial Distribution
- In Commercial Distribution
- Sterilization Method
- Ethylene Oxide
- Size
- Length — 5.5 InchWidth — 3.3 Inch
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Model / Reference | 610-1018 | 610-1018, 610-1031, 610-1124 |
| Device Class | Class IIa | Class II |
Other devices from Civco Medical Instruments Co., Inc.
Properties
Other properties (17)
- Implantable: No
- Reusable: No
- Contains Latex: No
- Reprocessed: No
- CMR Substance: No
- Human-derived: No
- Human Tissues: No
- Animal Tissues: No
- Endocrine Disruptor: No
- Administering Medicine: No
- Measuring Function: No
- Incorporates Medicinal Product: No
- Over-the-Counter: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No