Ascent Knee System 179290
Listed
Valid until
Not applicable
FDA recalls
None ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00880304263857 | Date24 Oct 2015 | StatusListed |
Recalls
-
- Terminated Z-0938-2019
Vanguard Knee System Series-A Standard Patella a potential commingle that could result in the size of the implant not matching the size on the label.
-
- Terminated Z-2289-2018
Two lots of tibial bearings were commingled. There is a possibility that a 14mm tibial bearing is packaged in a box labeled as 12 mm tibial bearing and vice versa. This may lead to extension of surgery to find another implant.
-
- Terminated Z-2177-2018
One lot of 75mm tibial tray is potentially etched and labeled as a 79 mm tibial tray. This may lead to delay in surgery, or immediate revision surgery to remove an incorrectly sized tibial component after it has been cemented.
-
- Terminated Z-2183-2018
The label master file was errantly set up as a 63/37mm instead of a 63/67mm. There is no 63/37 size offered, and the product is laser marked with the correct size.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Trade Name
- Ascent Knee System
- Generic Name
- Tibial insert
- Model / Ref / Catalog
- 179290
- Primary DI / UDI-DI
- 00880304263857
- FDA Product Code
- JWH, OIY, MBV, MBH, HRY
- GMDN Term
- Tibial insert
- Device Class
- Class II
- Regulation Number
- 888.3560, 888.3565, 888.3530
- GUDID distribution status
- In Commercial Distribution
- MRI Safety
- MR Conditional
Other devices from Biomet Orthopedics, Inc.
Properties
Other properties (8)
- Contains Latex: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Manufacturing Date: No