Skip to content

Bronchoscope guide 10000307

by Noah Medical Corp.

Bronchoscope guide 10000307 is a flexible video bronchoscope (FDA Class II; FDA product code EOQ, Bronchoscope (Flexible Or Rigid)) made by Noah Medical Corp., cleared under FDA 510(k) K223144 on 1 Mar 2023 and listed in FDA GUDID since 11 Dec 2024 (UDI-DI 00850048825031), active as of 22 Oct 2026. 5 FDA recalls are linked to it.

Active United States. Also listed in GUDID (US).

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

5 ongoing

This device is named in the recall

Documents

510(k) K223144

FDA submission, from GUDID

Trust

Unverified

Source GUDID, synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number00850048825031 UDI-DI Date11 Dec 2024ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE from GUDID Substantially Equivalent · Traditional NumberK223144 Date1 Mar 2023DecidedStatusActive

Official records

From the FDA record

Device description
Scope guide
MRI safety
MR Safe
Natural rubber latex
Labelled as containing natural rubber latex
Sterility
Supplied sterile; sterilise before use (Ethylene Oxide)

From the GUDID record (AccessGUDID), as read on 22 Oct 2026, as the labeller entered it.

Across registries

2 registries list this device

  • FDA 510(k) K223144 decided 1 Mar 2023 Active
  • FDA GUDID UDI-DI 00850048825031 listed 11 Dec 2024 Listed

Recalls

5 recalls · 5 ongoing · FDA

    • Ongoing Z-0598-2026 FDA Class II initiated 12 Sep 2025

    Due to Users not having access to the Instructions for Use/User Manual due to it being password protected.

    • Ongoing Z-0742-2025 FDA Class II initiated 28 Oct 2024

    Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of the bronchoscope when the system is not functioning can increase the chance of injury to the tissues of the lung as the scope is withdrawn including the possibility of pneumothorax.

    • Ongoing Z-0693-2025 FDA Class II initiated 25 Oct 2024

    Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal battery power; if during procedure, could cause interrupted/aborted procedure, may necessitate manual removal, which could lead to lung injury/pneumothorax.

    • Ongoing Z-0679-2025 FDA Class II initiated 25 Oct 2024

    Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive force may result in tissue injury/pneumothorax; instruments with long rigid distal sections may exhibit insertion resistance when bronchoscope bending section significantly articulated.

    • Ongoing Z-1734-2024 FDA Class II initiated 9 Mar 2024

    Due to a misaligned calibration of the wired controller that may result in unintended motion of the scope.

Timeline

7 dated events

  1. Cleared FDA 510(k) K223144
  2. Recall initiated FDA recalls Z-1734-2024
  3. Recall initiated FDA recalls Z-0693-2025
  4. Recall initiated FDA recalls Z-0679-2025
  5. Recall initiated FDA recalls Z-0742-2025
  6. Listed FDA GUDID 00850048825031
  7. Recall initiated FDA recalls Z-0598-2026

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

GUDID

Trade name
Bronchoscope guide
Generic name
Flexible video bronchoscope, reusable
Model / reference
10000307
Primary DI / UDI-DI
00850048825031
FDA product code
EOQ
GMDN term
Flexible video bronchoscope, reusable
Device class
Class II
Regulation number
874.4680
GUDID distribution status
In Commercial Distribution
Sterilisation method
Ethylene Oxide
MRI safety
MR Safe

Similar devices in this category

Other devices filed under Flexible video bronchoscope, reusable.

Other devices from Noah Medical Corp.

See all devices from Noah Medical Corp. →

Properties

  • Supplied sterile
  • Sterilise before use
  • Single use
  • Contains latex
  • Prescription use
  • Lot / batch number
  • Serial number
  • Expiration date
  • Manufacturing date
Other properties (6, all answered No)