Skip to content

Galaxy System

by Noah Medical Corp.

510(k) K223144

Galaxy System is a medical device (FDA Class II; FDA product code EOQ, Bronchoscope (Flexible Or Rigid)) made by Noah Medical Corp., cleared under FDA 510(k) K223144 on 1 Mar 2023, active as of 22 Oct 2026. 5 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

5 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK223144 Date1 Mar 2023ClearedStatusActive

Official records

From the FDA record

Received by the FDA
4 Oct 2022
FDA decision
1 Mar 2023 (Substantially Equivalent)
FDA review time
148 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K223144 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

5 recalls · 5 ongoing · FDA

    • Ongoing Z-0598-2026 FDA Class II initiated 12 Sep 2025

    Due to Users not having access to the Instructions for Use/User Manual due to it being password protected.

    • Ongoing Z-0742-2025 FDA Class II initiated 28 Oct 2024

    Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of the bronchoscope when the system is not functioning can increase the chance of injury to the tissues of the lung as the scope is withdrawn including the possibility of pneumothorax.

    • Ongoing Z-0693-2025 FDA Class II initiated 25 Oct 2024

    Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal battery power; if during procedure, could cause interrupted/aborted procedure, may necessitate manual removal, which could lead to lung injury/pneumothorax.

    • Ongoing Z-0679-2025 FDA Class II initiated 25 Oct 2024

    Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive force may result in tissue injury/pneumothorax; instruments with long rigid distal sections may exhibit insertion resistance when bronchoscope bending section significantly articulated.

    • Ongoing Z-1734-2024 FDA Class II initiated 9 Mar 2024

    Due to a misaligned calibration of the wired controller that may result in unintended motion of the scope.

Timeline

6 dated events

  1. Cleared FDA 510(k) K223144
  2. Recall initiated FDA recalls Z-1734-2024
  3. Recall initiated FDA recalls Z-0693-2025
  4. Recall initiated FDA recalls Z-0679-2025
  5. Recall initiated FDA recalls Z-0742-2025
  6. Recall initiated FDA recalls Z-0598-2026

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
EOQ
Device class
Class II
Regulation
21 CFR Part 874.4680
Advisory committee
Ear, Nose, Throat
FDA decision
SESE Substantially Equivalent

Devices cleared under K223144

6 devices · FDA GUDID

Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.

Devices cleared under K223144
UDI-DISize and differences
Accessory Kit 10000799
Noah Medical Corp.
1000079900850048825055
Prescription: no; Single use: yes
On the market
Bronchoscope 10000769
Noah Medical Corp.
1000076900850048825000
Prescription: yes; Single use: yes
On the market
Bronchoscope connector 10000339
Noah Medical Corp.
1000033900850048825017
Prescription: yes; Single use: yes
On the market
Bronchoscope guide 10000307
Noah Medical Corp.
1000030700850048825031
Prescription: yes; Single use: yes
On the market
Galaxy II system Wired Controller 10000111
Noah Medical Corp.
1000011100850048825062
Prescription: yes; Single use: yes
No longer on the market
Galaxy System GAL-001
Noah Medical Corp.
GAL-00100850048825048
Prescription: yes; Single use: no
On the market

Similar devices

Other devices filed under FDA product code EOQ (Bronchoscope (Flexible Or Rigid)).

Other devices from Noah Medical Corp.

See all devices from Noah Medical Corp. →