Cios Alpha
by Siemens Medical Systems USA, Inc.
510(k) K181560
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK181560 | Date24 Oct 2018 | StatusActive |
Recalls
-
- Terminated Z-2887-2020
Siemens has become aware of a hardware error which affects Cios Spin and Cios Alpha systems after the previous repair of the X-ray Generator with an integrated Energy Storage Unit (ESU).
-
- Terminated Z-2848-2020
Hardware error- Hot Plugging feature could cause an electrical malfunction on the main control D80 (below revision 04 installed) of the generator which is part of the main unit (c-arm) and cause permanent loss of imaging functionality.
-
- Terminated Z-1958-2020
- Terminated Z-1957-2020
Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the main cable can cause an unsafe electrical voltage (125 V DC) at the X10 connector and result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. These contact pins are accessible when the main cable is unplugged from the X10 connector
-
- Terminated Z-2127-2020
The unplugging of the main cable on the monitor trolley from the X10 connector can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector. This can result in an electrical shock if the user or other persons touch the contact pins of the X10 connector.
-
- Terminated Z-1953-2019
DAP chamber may be missing an insulating foil
-
- Terminated Z-1295-2019
Continuous operation of the Cios Alpha at high tube outputs may result in increased wear of radiation emitting components, risk of premature failure of the Monoblock and loss of imaging X-rays exists
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- OWB
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1650
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent