CRW® Uchrcbs
CRW Uchrcbs is a stereotactic radiosurgical invasive head frame (FDA Class II; FDA product code HAW, Neurological stereotaxic Instrument) made by Integra Lifesciences Corp., cleared under FDA 510(k) K160811 on 6 Dec 2016 (1 earlier submission) and listed in FDA GUDID since 14 Apr 2017 (UDI-DI 10381780252412), active as of 23 Oct 2026. 4 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number10381780252412 | Date14 Apr 2017 | StatusListed |
| United States | FDA 510(k)SESE | NumberK160811 | Date6 Dec 2016 | StatusActive |
Official records
- AccessGUDID record 10381780252412
- FDA 510(k) database K160811
- FDA 510(k) database K944463
From the FDA record
- Device description
- Universal Compact Head Ring Crossbar Attachment Screw
- MRI safety
- MR Conditional
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile; sterilise before use (Ethylene Oxide)
Across registries
- FDA 510(k) K160811 decided 6 Dec 2016 Active
- FDA GUDID UDI-DI 10381780252412 listed 14 Apr 2017 Listed
Recalls
-
- Terminated Z-0025-2016
Integra identified that the UCHRAP component (Arc Adapter Plate) would not assemble properly to the UCHRAR component (ARC Adapter Ring) of the Compact Head Ring Adapter due to an error in the manufacturing drawing.
-
- Terminated Z-2428-2012
- Terminated Z-2429-2012
T-handle screw is used on a complementary product to the CRW, the Integra HRAIM, which is the CT-compatible intubation Head Ring Assembly. When the T-handle on the HRAIM intubation hoop is tightened, the T-handle could stop in a vertical position that prevents the CRWPRECISE and certain models of the CRW-ASL from seating properly on the head ring although it appears to be fully seated
-
- Terminated Z-1400-2011
Overall length of the intubation hoop in the HRAIM Intubation Head Ring Assembly is too long and will not allow a device to to attach
Timeline
- Cleared FDA 510(k) K944463
- Recall initiated FDA recalls Z-1400-2011
- Recall initiated FDA recalls Z-2428-2012
- Recall initiated FDA recalls Z-2429-2012
- Recall initiated FDA recalls Z-0025-2016
- Cleared FDA 510(k) K160811
- Listed FDA GUDID 10381780252412
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Crw®
- Generic name
- Stereotactic radiosurgical invasive head frame
- Model / Ref / Catalog
- UCHRCBS
- Primary DI / UDI-DI
- 10381780252412
- FDA product code
- HAW
- GMDN term
- Stereotactic radiosurgical invasive head frame
- Device class
- Class II
- Regulation number
- 882.4560
- GUDID distribution status
- In Commercial Distribution
- Sterilisation method
- Ethylene Oxide
- MRI safety
- MR Conditional
Similar devices
Other devices filed under FDA product code HAW (Neurological stereotaxic Instrument).
- 9733977 by Medtronic Navigation Inc.
- Navient Accessory kit - cranial 955-NC-AKC by ClaroNav Kolahi Inc
- Stryker Power Tool Navigator Attachment by Stryker Corp.
- 9730092 by Medtronic Navigation Inc.
- Navigus Passive Probe (BL), Model NP-2000 by Image-Guided Neurologics, Inc.
- ATLASPLAN by Radionics, Inc.
- SOLOPASS 2.0 System by Intravent Medical Partners, LP
- NEXDRIVE™ w/DIG MI-1000-00771 by FHC, INC.
- 6000-055-000 by Stryker Leibinger
- Viewpoint Tools Sterrad by Philips Medical Systems (Cleveland), Inc.